Through-flow control of the internal heater

V4
Upper limit value (Tih)
n
Degassing mode
n
Fill mode
n
Cont.outp.limit
n
T_triac
n
The relevant versions of the installed software and hardware are required for
on-site service visits and telephone support.
1.
In the Device status menu select the menu item  SW version or
 Hardware version.
The software versions installed on the device are displayed. If
available, the software versions of any connected interface mod-
ules are also displayed. The hardware version is not displayed for
all modules. In the case of older modules, no hardware version
can be recorded by software and displayed. In such cases "---" is
displayed.
The device type is displayed in the Device type menu item.
The serial numbers are required for on-site service visits and telephone
support.
1.
Select the  Serial numbers menu item in the Device Status menu.
The serial numbers of the device components are displayed.
This section is relevant for:
Integral XT devices
n
The bypass valve is used to ensure that there is sufficient flow
through the heater in the device to prevent damage to the heat
transfer liquid.
1.
If the flow rate in the device is too low, a sound is heard for three
seconds.
The heating switches off on all poles, the pump and the cooling
unit are switched off.
2.
In the display the message appears that the flow is too low.
3.
Eliminate the cause.
If the reason for the low flow rate is the high flow resistance caused
by the application, turn the adjusting wheel of the bypass valve coun-
terclockwise. Turn the adjusting wheel so far until the required flow is
reached.
4.
Press the Enter key.
Also press the Enter key if the device was switched off in the fault
state. Warnings disappear automatically when the cause is eliminated.