Obligations Of The Operator - Lauda IN 550 XT FC Operation Manual

Integral process thermostats with flow control unit
Hide thumbs Also See for IN 550 XT FC:
Table of Contents

Advertisement

1.2

Obligations of the operator

V4
It is essential to avoid the penetration of secondary liquids into the
n
device (e.g. via a defective heat exchanger provided by the customer).
Observe all product safety labels/safety marks on the device.
n
Additional safety information for water-cooled devices
Use suitable cooling water to prevent corrosion in the cooling water
n
circuit.
Secure the water cooling return hose in the outlet area to prevent the
n
hose from jerking suddenly, even when pressure surges occur.
The return hose of the water cooling system must be fixed in the outlet
n
area in such a way that hot cooling water cannot splash out.
Avoid bending or squeezing the return hose of the water cooling system.
n
Gage pressure can cause the cooling water hoses to rupture and hot
cooling water to escape.
In order to avoid damage caused by leakage of the cooling water system,
n
it is recommended to install a leakage water detector with water shut-
off.
The national regulations for operation applicable in the country in which the
system is installed must be complied with.
In particular, the application of statutory regulations concerning operational
safety must be observed.
Note the installation conditions outlined in Ä "Installation location"
on page 149.
Operators within the EU must meet the applicable provisions of Regulation
(EU) No. 517/2014 on fluorinated greenhouse gases. The regulation provides
a complete overview, and includes:
The general purpose of the regulation is to reduce emissions of fluori-
n
nated greenhouse gases.
Regular tightness checks, depending on the quantity of CO
n
(see type plate and regulation; regular tightness checks may be neces-
sary if the CO
Having leakage inspections, maintenance, repairs, decommissioning
n
or recovery work carried out by certified, authorized personnel (for
example LAUDA Service).
Keeping records of refrigerants added or recovered, including quantity
n
and type. Records must be kept for a minimum of 5 years.
Integral process thermostats with flow control unit
equivalent exceeds 5 t).
2
equivalent
2
9 / 171