12 Calibration service

12 Calibration service
The ISO 9001 and GLP guidelines require regular inspection of your vol-
ume measuring devices. We recommend performing a volume check ev-
ery 3 to 12 months. The cycle is dependent on the individual requirements
of the device. Checks should be performed more frequently, in case of
high frequency of use or the use of aggressive media.
The detailed testing instructions are available for download on
www.brand.de or www.brandtech.com.
BRAND also offers the possibility to have your instruments calibrated by
our factory calibration service or by the BRAND DAkkS laboratory.
Simply send us the instruments to be calibrated, accompanied by details
about which type of calibration you would like. The instruments will be
returned to you after a few days together with a test report (factory cali-
bration) or a DAkkS calibration certificate. More information can be ob-
tained from your dealer or directly from BRAND.
The order document is available for download on www.brand.de (see
Technical Documents).
For customers outside Germany
If you would like to use our calibration service, please contact one of our
service partners in your region. If a factory calibration is requested, the
service partner can forward the instrument to BRAND.