Safety, Care, Accuracy Test And Disposal - Microlife BPA7 Touch BT Manual

Bluetooth blood pressure monitor
Hide thumbs Also See for BPA7 Touch BT:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 3
Error
Description Potential cause and remedy
«Err 2»
Error signal
During the measurement, error signals
were detected by the cuff, caused for
BK
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«Err 3»
Abnormal
An adequate pressure cannot be
cuff
generated in the cuff. A leak may have
BT
pressure
occurred. Check that the cuff is
correctly connected and is not too
loose. Replace the batteries if neces-
sary. Repeat the measurement.
«Err 5» Abnormal
The measuring signals are inaccurate
result
and no result can therefore be
displayed. Read through the checklist
for taking a reliable measurement and
then repeat the measurement.*
«Err 6» AFIB/MAM
There were too many errors during the
Mode
measurement in AFIB/MAM mode,
making it impossible to obtain a final
result. Read through the checklist for
taking a reliable measurement and
then repeat the measurement.*
«HI»
Pulse or cuff
The pressure in the cuff is too high
pressure too
(over 299 mmHg) OR the pulse is too
high
high (over 200 beats per minute).
Relax for 5 minutes and repeat the
measurement.*
«LO»
Pulse too low The pulse is too low (less than
40 beats per minute). Repeat the
measurement.*
Problem with
If any problem occurs with the Blue-
®
Bluetooth
tooth connection, the Bluetooth
connection
BS blinks rapidly for approximately 10
seconds. To solve the problem, please
visit www.microlife.com/connect.
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
Microlife BP A7 Touch BT

11.Safety, care, accuracy test and disposal

Safety and protection
 Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
 This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
 This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section.
 The cuffs are sensitive and must be handled with care.
 Only pump up the cuff once fitted.
 Do not use this device if you think it is damaged or notice
anything unusual.
 Never open this device.
 Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
 The measurement results given by this device is not a diag-
nosis. It is not replacing the need for the consultation of a physi-
cian, especially if not matching the patient's symptoms. Do not
rely on the measurement result only, always consider other
potentially occurring symptoms and the patient's feedback.
Calling a doctor or an ambulance is advised if needed.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.
Contra-indications
Do not use this device if the patient's condition meets the following
®
icon
contra-indications, to avoid inaccurate measurements or injuries.
 The device is not intended for measuring blood pressure in
pediatric patients of age younger than 12 years old (children,
infant, or neonates).
 Presence of significant cardiac arrhythmia during measurement
may interfere with blood pressure measurement and affect the
reliability of blood pressure readings. Consult with your doctor
about whether the device is suitable for use in this case.
7
EN

Advertisement

Table of Contents
loading

Table of Contents