Thermo Scientific Nicolet iS50 User Manual page 27

Hide thumbs Also See for Nicolet iS50:

Advertisement

Cleaning the Sampling Plate
Thermo Scientific
You can remove the sampling plate if necessary for cleaning.
To remove the sampling plate, raise the pressure arm by turning the control knob
counterclockwise and then lift the plate off the assembly.
Use mild soap and water or, if needed, alcohol, toluene or a similar organic solvent to clean
the sampling plate. The plate can be disinfected with a mild bleach solution.
NOTICE Prolonged exposure to strong acids or bases may stain the sampling plate.
When the sampling plate is clean, install it on the ATR module. See
the Sampling
Plate" for more information.
Pressure control knob
Pressure arm
Sampling plate
"Removing and Installing
iS50 ATR Module User Guide
1
iS50 ATR Module
25