Co Calibration Procedure - Thermo Scientific BB 150 User Manual

Co2 incubator
Table of Contents

Advertisement

10.
Maintenance
10.6
CO
calibration procedure
2
1. Activate the calibration:
Keep the
All operating panel indicators flash.
2. Display the nominal value:
Press the
The set nominal value of 5 % is displayed.
3. Enter the measured value (destination value):
Press the
or
Press the
Destination value e.g. 5.6 %.
4. Accept the destination value:
Press the
The CO
display momentarily shows "CAL",
2
then the corrected actual value (measured destinati-
on value 5.6 %) is displayed.
5. Cancel the calibration process:
Press any key.
The temperature display and the CO
the actual values.
key depressed for 5 seconds.
key.
+
keys.
+
keys.
key.
display show
2
CO
Incubator BB 150 Operating Instructions