Cleaning and disinfection
Heating phase: Remaining run time approx 25 hours
The work space is heated to a temperature of 90° C while an elevated relative
humidity is created. The current decontamination temperature is shown at the
temperature display.
Decontamination phase: Remaining run time approx 23 hours
After the decontamination atmosphere has been created, the decontamination
phase with a run time of 9 hours is started.
If the door is opened during this time, the decontamination routine is restarted
automatically as soon as the door has been closed.
Cool-down phase: Remaining run time approx 14 hours
The device cools down until the originally set temperature nominal value is reached.
Postheating phase: Remaining run time approx 3 hours
During the postheating phase, condensate within the device is eliminated as far
as possible; remaining condensate accumulates at the bottom of the work space.
End of the decontamination routine: Remaining run time 0 hours
When the remaining run time has elapsed to 0 hours, the device has reached the
originally set working temperature again (e.g. 37° C). The ContraCon
decontamination routine must then be ended by pressing the appropriate key.
NOTE – Overtemperature
If the maximum temperature of 95° C is
exceeded
decontamination routine, the routine is
interrupted and the device heating is switched
off.
Fig. 22
ContraCon decontamination routine
procedure
during
the
ContraCon
CO
Incubator BB 150 Operating Instructions
2
51