Temperature Calibration Procedure - Thermo Scientific BB 150 User Manual

Co2 incubator
Table of Contents

Advertisement

10.
Maintenance
10.4

Temperature calibration procedure

1. Activate calibration:
Keep the
All operating panel indicators flash.
2. Display the nominal value:
Press the
The preset value of 37° C is displayed.
3. Enter the measured value (destination value):
Press the
or
Press the
Destination value e.g. 36.4° C
4. Accept the destination value:
Press the
The temperature display momentarily shows "CAL",
then the corrected actual value (measured destination value 36.4° C)
is displayed.
5. Cancel the calibration process:
Press any key.
The temperature display and the CO
key depressed for 5 seconds.
key.
+
keys.
+
keys.
key.
display show the actual values.
2
CO
Incubator BB 150 Operating Instructions