Tonic T20X1 Readhead Installation Drawing - Renishaw TONiC T20 1 Series Installation Manual

Resm angle encoder system
Hide thumbs Also See for TONiC T20 1 Series:
Table of Contents

Advertisement

TONiC T20x1 readhead installation drawing

2 off mounting holes M2.5 through,
counterbored Ø3 x 2.75 deep from
alternative mounting face
R > 20 Dynamic bend radius
R > 10 Static bend radius
4.6
*
Extent of mounting faces.
UHV cable 3.0 approx.
Dynamic bend radius not applicable for UHV cables.
TONiC RESM20 angle encoder system installation guide
Optical centreline (incremental and reference mark)
'A' section ring – Offset 1.75 ±0.5
18
'B' section ring – Offset 3.25 ±0.5
31
16
3
22
29
6 min
35
23
11.5
(Yaw tol. ±0.4°)
7.5
3
Set-up LED
CAL/AGC LED
2 off mounting holes M2.5 through, counterbored Ø3 x 2.3 deep both sides
NOTE: The recommended thread engagement is 5 min (7.5 including counterbore)
and the recommended tightening torque is between 0.25 and 0.4 Nm.
Rideheight 2.1 ±0.15
Forward direction
of ring (increasing count)
(Roll tol. ±0.5°)
0.25
0.08
11 *
13.5
4.25
Alternative
mounting face
Recommended
mounting faces
4.15
(Pitch tol. ±1°)
0.6
Dimensions and tolerances in mm
8.75 *
10
12.1