Service and maintenance

8.3 Keying in sensor data

8.3
Keying in sensor data
The following step-by-step procedure only applies to:
• FUS060 transmitters with SONOKIT sensors
• Replacement transmitters
NOTICE
SONO 3100 and SONO 3300
All FUS060 transmitters calibrated with sensors SONO 3100 or SONO 3300 are delivered with
sensor data preset from factory. The data should not be changed
The step-by-step procedure describes the required parameters/data to be entered and the
sequence of entering them in order to configure the transmitter correctly for a given pipe
application.
Note
The pipe data are taken from the SONOKIT sensor geometry measurement report or from
another sensor data sheet or calibration report.
Entering the data
1. Enter the sensor diameter (Di) (menu 7.1.3.1).
Note
If Di is smaller than registered, the setting in step 9 (displacement) must be changed first.
2. Choose appropriate engineering unit for flow (menu 3.1.1).
3. Enter maximum flow (menu 3.1.2).
4. Choose "AUTO" for calibration choice (menu 7.1.1).
5. Enter viscosity for media (menu 7.1.3.1), otherwise the default value (0.01 cm
20 °C) will be used.
6. Enter inside roughness of the sensor pipe (menu 7.1.3.2), otherwise the default value (0.4
mm) will be used.
7. Enter the measured transducer distance for each path (menus 7.1.4.x.1).
8. Enter the measured angle for each path (menus 7.1.4.x.2).
9. Enter the measured displacement for each path (menu 7.1.4.x.3). Enter 0 for paths not used.
10.Check the number of paths (menu 7.2). They are preset from factory according to order.
68
SITRANS FUS060 with HART
Operating Instructions, 09/2021, A5E01204521-AL
/s for water