Technical specifications
10.1 SITRANS FUS060
Output
Configuration conditions
Configuration
Passive – Pos
Passive – Neg
Passive – Pos
Passive – Neg
Passive – Pos
Passive – Neg
Active – Pos
Active – Neg
Digital output 2 (relay)
Relay, NC or NO contact
For explosion protection
Output function, configura‐
ble for:
HART communication via ana‐
log output (4 to 20 mA)
Load
Cable
Protocol
Electrical isolation
Table 10-3
Accuracy
Accuracy *
Pulse output
Analog output
Repeatability
Reference conditions (water)
Process temperature in the
connected sensor
Ambient temperature at the
transmitter
76
Supply
Load
12 V
75 Ω
12 V
75 Ω
12 V
520 Ω
12 V
520 Ω
12 V
6.1 kΩ
12 V
6.1 kΩ
Internal (25 V)
1 kΩ
Internal (25 V)
1 kΩ
Switching capacity max. 5 W
Max. 50 V DC, max. 200 mA (DC)
Self-resetting fuse, Ri = 9 Ω
Max. 30 V DC, max. 100 mA (DC), max. 50 mA (AC); cf. ATEX approval certificate
Alarm, flow direction, limits, no function
Via PC/laptop (PDM) or HART communicator
≥ 230 Ω
2-wire shielded ≤ 3 km (≤ 1.86 miles)
Multi-core shielded ≤ 1.5 km (≤ 0.93 miles)
HART, version 5.1
Outputs electrically isolated from power supply and from each other (output isolation 63 V
AC / 100 V DC to ground)
≤ ±0.5% of measured value at 0.5 to 9 m/s or
≤ ±0.25 / V [m/s] % of measured value at flow < 0.5 m/s
As pulse output plus ±0.1% of measured value ±20 µA
≤ ±0.25% of measured value at 0.5 to 9 m/s
25 °C ±5 °C (77 °F ±9 °F)
25 °C ±5 °C (77 °F ±9 °F)
Current
Output voltage limits
(max./min)
160 mA
12 V / 600 mV
160 mA
12 V / 600 mV
23 mA
12 V / 100 mV
23 mA
12 V / 100 mV
2 mA
12 V / 0 mV
2 mA
12 V / 0 mV
21 mA
21 V / -
21 mA
21 V / -
SITRANS FUS060 with HART
Operating Instructions, 09/2021, A5E01204521-AL