Siemens SITRANS FUS060 Operating Instructions Manual page 77

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS FUS060:
Table of Contents

Advertisement

Accuracy *
Installation conditions of
connected sensor
Transmitter warming-up
time
* Typically depending on accuracy of installation measurement
Table 10-4
Rated operation conditions
Rated operation conditions
Ambient conditions
Ambient temperature
Operation
In potentially explosive at‐
mospheres
Storage
Enclosure rating
Electromagnetic compatibility
Emission
Immunity
Medium conditions
Processs temperature
Gases/solids
Table 10-5
Design
Design
Transmitter (only as remote version)
Enclosure material
Wall mounting bracket (standard and
special)
Weight
Cable glands
SITRANS FUS060 with HART
Operating Instructions, 09/2021, A5E01204521-AL
Upstream section > 10 x DN and downstream section > 5 x DN
30 minutes
-20 to +50 °C (-4 to +122 °F)
Observe temperature classes
-25 to +80 °C (-13 to +176 °F)
IP65 / NEMA 4
For use in industrial environments
EN 55011/CISPR-11
EN 61326-1 (Industry); use of shielded output cables is recommended
The measuring media must be ultrasonic signal compatible. It must be homogene‐
ous and not two-phased to transfer the acoustic ultrasonic signals
-200 to +250 °C (-328 to +482 °F) (not directly influenced by medium temperature)
Influence accuracy of measurement (approx. max. 3 % gases or solids)
Transmitter is connected to the transducers via 3 to 120 m (9.8 to 395 ft) long spe‐
cially shielded cables (coaxial cable).
For ATEX versions mounted in the Ex area only with 3 m (9.8 ft) long cables
Die-cast aluminum, painted
Stainless steel (standard: always incl.)
4.4 kg (9.7 lb)
Power supply and outputs: 2 x M20 or 2 x ½" NPT
Transducers/sensor: 2/4 x M16 or 2/4 x ½" NPT
Technical specifications
10.1 SITRANS FUS060
77