13.1.4.3 Tube head seasoning
About this task
Steps
User's manual
5. Press and hold down the exposure button for the duration of the
exposure.
The C-arm does not move when you take a test exposure.
6. Select the green check mark button.
Follow the steps below to perform tube head seasoning.
1. Select User > 1500 Testing Routines > 1530 Tube Head Seasoning.
This option allows you to warm up the X-ray tube, i.e. run a tube head
seasoning process. This is necessary if the X-ray unit has not been used
for a week or more and/or if you receive error message E332 (Severe
arcing across X-ray tube).
2. Select the START button.
3. Move to a protected area.
4. Press the exposure button when the word Ready appears.
You can press and hold down the exposure button for the whole duration
of the process or lift your thumb from the exposure button when the word
Wait appears.
NOTE
The seasoning process takes several minutes.
After a successful process the message OK displays.