Indication
Problem
Cooler and gas
Middle LED on
pump in
cooler is green;
operation;
condensate in
sample gas lead;
LED of the LA
electronics is
green;
17

PROPER DISPOSAL OF THE DEVICE

At the end of the service life of our products, it is important to take care of the appropriate disposal of obsolete
electrical and non-electrical devices. To help protect our environment, follow the rules and regulations of your
country regarding recycling and waste management.
www.mc-techgroup.com
Possible Cause
Pre-filter clogged up;
Pump hose faulty;
Peristaltic pump SR25.1
faulty;
Not sufficient drying of
sample gas;
Optional liquid alarm
sensor:
Sensor has not turned
pump off.
PSS-10 | 1.00.02
Check/Solution
Remove filter from condensate lead;
Pump delivering?
Change filter;
Pump not delivering?
Change pump hose (see manual SR25.1);
OK?
Check peristaltic pump (see manual SR25.1);
OK?
Check cooler (see manual ECP20-1);
Check the LA electronics and if necessary replace.