INSTALLATION INSTRUCTIONS

C A U T I O N !
When in use, the gas conditioning system case should be placed in an area well
away from any heat emitting sources in order to prevent damage caused by an
accumulation of heat. The air vents in the case must be free at all times.
N O T E !
In order to ensure that the case is safe and stable when being used, it should be
laced on a horizontal surface.
The equipment is to be used in a vertical position only. The perfect functioning of
the separation and drainage procedures will only be guaranteed if the equipment is
used in a vertical position.
When the equipment is being used outside, ample protection against the effects of
direct sunlight and dampness must be provided. In winter, the equipment must
only be used in frost-free areas.
It is recommended that the case be protected using one of the many models
available.
In order to guarantee the operational safety of the gas conditioning system and the
additionally connected analysers, and to avoid false alarms, the gas conditioning
system should not be used at temperatures other than those specified.
It is of great importance that the analysers which have been additionally connected
be used at temperatures well above the specified gas outlet dew point of +5 °C [41
°F]. This prevents the gas in the connector lines from condensing completely.
In the event of the unheated sample gas lines being connected to the gas
conditioning system on a slope, it is not necessary to carry out a preliminary
condensate removal.
www.mc-techgroup.com
Heavy device!
Risk of injury when handling heavy equipment.
Do not lift, move or carry the device without help. A second person is
required to lift, move or carry the device.
PSS-10 | 1.00.02