Service and troubleshooting
September 2022
WARNING
Process leaks could result in death or serious injury.
Ensure that the transmitter is handled carefully. If the process seal is damaged, gas might
escape from the tank.
7.2

Diagnostic messages per NAMUR NE 107

7.2.1
Electronics failure, transmitter
Category
Cause
An electronics error has occurred.
The device measurement reading is invalid.
Recommended actions
Related information
Restart the device
7.2.2
Electronics failure, sensor module
Category
Cause
An electronics error has occurred.
The device measurement reading is invalid.
Recommended actions
Related information
Restart the device
56
Failure
1. Restart the device.
2. If the condition persists, replace the transmitter electronics.
Failure
1. Restart the device.
2. If the condition persists, replace the transmitter electronics (or the device).