Service and troubleshooting
September 2022
Dropping of level close to tank bottom
Symptom
Measured value drops to zero level in the tank bottom region.
Figure 7-11: Symptom
A. Level
B. Time
C. Actual level
D. Reported level
Possible cause
Transmitter has locked on a strong tank bottom echo.
Recommended actions
Related information
Reference height
Use tank bottom projection
Enable bottom echo visible when tank is empty
80
A
C
D
Verify the Reference Height is configured correctly.
Enable the Tank Bottom Projection function
Enable the Bottom echo visible when tank is empty parameter.