Emerson Rosemount 3408 Reference Manual page 97

Level transmitter, non-contacting radar
Hide thumbs Also See for Rosemount 3408:
Table of Contents

Advertisement

Reference Manual
00809-0100-4418
Only suppress echoes which can be clearly identified as objects in the tank. Compare
the list of interfering echoes with the tank drawing or visual inspection of the tank.
Do not suppress false echoes located below the product surface.
Keep the number of suppressed false echoes to a minimum.
It is recommended to suppress false echoes using Rosemount Radar Master Plus.
Procedure
1. Select Service Tools → Echo Curve.
2. In the echo curve, click
Postrequisites
It may be necessary to suppress new false echoes at a later stage when objects have
become visible due to surface movement.
Related information
Amplitude threshold
Delete a false echo suppression
Prerequisites
It is recommended to suppress false echoes using Rosemount Radar Master Plus.
Procedure
1. Select Service Tools → Echo Curve.
Rosemount 3408 Level Transmitter
Figure 7-21: Add False Echo Suppression
at the unknown echo peak, and then select Suppress.
Service and troubleshooting
September 2022
97

Advertisement

Table of Contents
loading

Table of Contents