Pin
Wire
Pin #
1
Name
Color
7
Ain1
violet
grey-
8
Ain2
pink
6
Ain_GND
blue
blue-
9
Aout1
red
white-
10
GND
green
shield
2
Dout1
brown
1
Dout2
white
5
Din1
pink
4
Din2
grey
3
Din_COM
yellow
11
Pout
red
12
NC
black
1: Wires of compatible cables: ICS-2.1-xx
PL-4063-00, Rev03, DCO# 20-275
Standard
Description
Designation
Analog Input
Actual Process
(Current)
Value
Analog Input
Reference
(Voltage)
Value
Analog Input
--
Ground
Actual Speed
Analog Output
Actual Process
Value
Ground
--
Digital Output
Status
Digital Output
Error
Enable
(Reset)
Process Mode
Common
--
Digital Input
Output
Supply
24 VDC
Output
Not connected
--
Table 12: PLC connector of EasyConnect driver
User Manual for PuraLev
Typical Levels
4..20 mA = 0..100%
0..10 V = 0..10000 rpm (Speed Mode)
-> Speed limit = 9000 rpm
-> Cut-off (min.) speed = 0 rpm
0..10 V = 0..100% (Process Mode)
--
0..10 V = 0..10000 rpm (Speed Mode)
0..10 V = 0..100% (Process Mode)
--
Closed circuit active, system on
Open circuit not active, system off
Closed circuit not active, system on
active, system off
Open circuit
5-24 V active
not active
0 V
5-24 V active
not active
0 V
--
24VDC ±10%, 200 mA
(Max. current is together with Pout of
Fieldbus OUT connector in Table 11)
--
2: Designations are for standard firmware.
®
i100SU
www.levitronix.com
Note
450 Ohm shunt input, no galvanic isolation.
The designation of the analog inputs can be
®
changed with Levitronix
Service Software.
Analog voltage input: 0 – 10V
(7.9 kOhm, no galvanic isolation).
Note: Max. input voltage of 11 V shall not be
exceeded.
The designation of the analog inputs can be
®
changed with Levitronix
Service Software.
Reference to Ain1/2.
Direct connection, no protection. Galvanic
isolation on user side is required. GND is
reference.
The configuration of the analog output can be
®
changed with Levitronix
Service Software.
Reference for Aout1, Dout1, Dout2 and Pout.
Open drain, max. 24V, 100mA
This signal indicates the state of the pump
system.
The configuration of the digital outputs can be
®
changed with Levitronix
Service Software.
Open drain, max. 24V, 100mA
When active, the system drives the impeller to
zero rpm and shuts down. With a reset pulse
(see "Enable" description) the system can be
re-initialized.
The configuration of the digital outputs can be
®
changed with Levitronix
Service Software.
Galvanic separation with optocoupler and
2.2 k input resistance.
The "Enable" signal switches the pump
system on and off. Resets pump from error
state with 300-700ms pulse.
Galvanic separation with optocoupler and
2.2 k input resistance.
Switch between process mode and speed
mode. In process mode the closed loop PI
controller can be configured with Levitronix
Service Software.
Reference to Din1 and Din2.
For supply of external devices (displays etc.).
Internally protected.
Reference is GND.
--
®
22