7.5
Test of signal converter using GS 8 A simulator (option)
7.5.1 IFC 020 K und IFC 020 F signal converter
GS 8 A operating elements and accessories
Y
20
20
10
40
5
80
2,5
160
1,25
320
Electrical connection
H2
Z
C 5
B
X 5
X 3
amplifier PCB, see
Sect. 8.8
a) Switch off power source before
opening the housing !
b) Unscrew the 4 recessed head screws, see
Fig. A in Sect. 8.1, and remove transparent
cover from the signal converter housing.
c) Unscrew recessed head screw, see Fig. B in
Sect. 8.1, and remove black plastic cover.
7/5
+
0
between GS 8 and IFC 020 K and IFC 020 F signal converter.
Power
L
Connection of milliammeter and
electronic frequency counter,
see Sect. 2.4 "Connection of outputs".
Installation and operating instructions IFC 020
P
D
L
H
Using GS 8 simulator
Additional adapter is required
Order No. 2.10764.00.
Outputs
N
I⊥
I
I+
10
d) Pull off blue 9-pin plug from the amplifier PCB,
see Sect. 8.9:
socket X3 field power supply and socket X5
signal cable.
e) Connect plug B to socket X3 (2-pin) and
plug C (5-pin) to socket X5 (5-pin)
B
Plug for field power
supply, 2-pin
C5
Plug for signal cable,
5-pin
D
Switch,
flow direction
H
Socket for H2 plug
on cable Z
H2
Plug of cable Z
L
Power supply ON
P
Potentiometer "zero"
X3
Socket for plug B
on amplifier PCB
X5
Socket for plug C3
on amplifier PCB
Y
Switch,
measuring ranges
Z
Cable between GS 8 A
and signal converters
IFC 020K and IFC 020 F
U-clamp terminal
P
P⊥ S
PE
Milliammeter, accuracy
A
class 0.1, R
i
range 4-20 mA
Electronic frequency
counter, input resistance
approx. 1 kΩ, range
0-1 kHz, time base min. 1
second, see connection
diagrams in Sect. 2.4.
< 500 ohms,