IFC 300 1.1 Scope of the document These instructions are supplementary to the standard handbook of the IFC 300 with ER 3.4.0 (refer to document MA IFC 300 R05). They provide additional information for the devices when being operated and connected to a PROFINET IO network.
This device is a digital device used exclusively as an electronic control system utilised by a public utility or in an industrial plant. INFORMATION! This PROFINET IO device is certified by the PROFIBUS Nutzerorganisation e.V. 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en www.krohne.com...
These instructions also contain detailed descriptions on how to use grounding rings and how to install the flow sensors in metal or plastic pipes or in pipes which are coated on the inside. 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en www.krohne.com...
Unused M12 connectors must be covered by a cap to protect the connector from dust and • humidity. Straight M12 connectors are recommended. • Standardised M12 connectors (IP66/67 or higher) must be used and are described in the following standards: IEC 61076-2-101 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en www.krohne.com...
3.4.3 Assembly of the M12 data connector with PROFINET IO Figure 3-2: PROFINET IO M12 data connector (female, D coded) Number Signal Colour yellow white orange blue Table 3-5: PROFINET IO M12 data connector pinout and colouring scheme www.krohne.com 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en...
Devices with two Ethernet connectors can still be used. The unused connectors must be covered by the provided caps (for details refer to Overview of the PROFINET IO M12 connectors on page Figure 3-3: Point-to-point or star topology 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en www.krohne.com...
The control system must support this topology. This topology is typically preferred if redundancy is required. INFORMATION! Devices in this network topology need two Ethernet ports. Figure 3-5: Ring topology www.krohne.com 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en...
4.1.2 Device discovery and identification (DCP signal) To identify the device, the DCP signal service can be used. While a DCP signal is active a correspondent message is shown on the device display status page (for details refer to IFC 300 handbook).
Standard PROFINET IO DAP with up to 2 ports 32768 Interface 32769 Port 1 32770 Port 2 Sensors Module representing the sensor system Sensor Data I/O data, start-up records Table 4-4: Static PROFINET IO slot/subslot mapping www.krohne.com 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en...
(UINT) or bit-wise (WORD) depending on context. Byte order on bus Significance Example Value = 1234 0x04 0xD2 Table 4-6: Data type Unsigned16 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en www.krohne.com...
Table 4-8: Description of I&M1 4.5.3 I&M2 - Installation date I&M2 parameters can be changed by user. Name Length Description INSTALLATION_DATE Date of installation or commissioning. Table 4-9: Description of I&M2 www.krohne.com 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en...
PROFINET IO connection is established. The following sections provide a list of all parameters within the record data. A unique parameter ID (ID) is introduced in order to reference each parameter explicitly. 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en www.krohne.com...
Unit: selected by ID 14 Empty Pipe Detection Unsigned8 Turn empty pipe detection on or off. Note: empty pipe detection is based on conductivity measurement. Table 4-13: Parameter group - Measurement settings www.krohne.com 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en...
Unit selector for flow velocity values. Level Unsigned16 Unit selector for level values. 1 Table 4-14: Parameter group - Unit settings 1 Only available for signal converters with a TIDALFLUX flow sensor. 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en www.krohne.com...
Low flow cut-off hysteresis if measurement (ID 24) is set to "Mass flow". Unit: selected by ID 8 Table 4-16: Parameter group - Totalizer 2 settings 1 Only available for signal converters with a TIDALFLUX flow sensor. www.krohne.com 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en...
The totalizer can be controlled by the corresponding control octet in the output data (for details Output data refer to on page 24). 4.8.3 Diagnosis blocks A diagnosis block provides a list of bits that give information about the internal device diagnosis. 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en www.krohne.com...
Conductivity – Status 1 Status 36...37 Unsigned16 For details refer to information on page 26. Units of 38...39 Conductivity – Unit 1 Unsigned16 For details refer to measurement on page 17. www.krohne.com 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en...
Page 23
Cyclic diagnosis (Octet 5) Unsigned8 Cyclic diagnosis (Octet 6) Unsigned8 Cyclic diagnosis (Octet 7) Unsigned8 Table 4-18: Input data 1 No valid values when used with CAPAFLUX flow sensor! 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en www.krohne.com...
This example assumes a constant flow that is being totalized. Figure 4-1: Example for totalizer control X: time Y1: totalizer value Y2: totalizer run/stop Y3: totalizer reset Value Description STOP RESET 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en www.krohne.com...
Initial value as long as no measurement is available. UNCERTAIN - process related 0x78 Process value is potentially invalid (accuracy not guaranteed). GOOD 0x80 Valid process value. 0x01 Simulation running. Table 4-20: Status coding www.krohne.com 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en...
NOTES IFC 300 09/2017 - 4005824401 - AD IFC 300 Profinet R01 en www.krohne.com...