Cleaning and disinfecting
Hide thumbs Also See for SonoSite S Series:

Advertisement

5 Clean the exterior surfaces of the ULTRASOUND SYSTEM to remove any debris or bodily fluids. Use the
following procedure:
a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner
from
Table 4-2, "Cleaners and disinfectants compatible with Edge system and
page 53.
b Remove all gel, debris, and bodily fluids from the system.
c With a new wipe, clean the system, including the display, by wiping from clean areas to the soiled areas.
This method helps to avoid cross-contamination.
d Observe the minimum wet contact time recommended by the manufacturer. Monitor the system for
wet appearance. Re-apply with a new wipe if no longer wet.
e Allow the ultrasound console to air dry in a clean, well-ventilated space.
6 Clean the TRANSDUCER CABLE AND BODY to remove any debris or bodily fluids. Use the following
procedure:
a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner
from
Table 4-2, "Cleaners and disinfectants compatible with Edge system and
page 53.
b Remove all gel, debris, and bodily fluids from the transducer.
c With a new wipe, clean the cable and transducer, starting from the cable, wiping toward the scanhead.
This method helps to avoid cross-contamination.
Caution
d Observe the minimum wet contact time recommended by the manufacturer. Monitor the transducer
for wet appearance. Re-apply with a new wipe if no longer wet.
7 Verify that all gel, debris, and bodily fluids have been removed from the system and transducer. If
necessary, repeat steps 5 & 6 with a new wipe.
WARNINGS
46
Do not allow moisture near the electronic components of the connector.
Failure to remove all gel, debris, and bodily fluid could leave contaminants on the
probe.
transducers"on
transducers"on
Cleaning and disinfecting