Cleaning and disinfecting
Hide thumbs Also See for SonoSite S Series:

Advertisement

8 Prepare the disinfectant for use.
a Choose a high-level disinfectant from
NanoMaxx
transducers"on page 142.
b Check the expiration date on the bottle to ensure the disinfectant has not expired.
c Mix or check that the disinfection chemicals have the concentration recommended by the manufacturer
(for example, a chemical strip test).
d Check that the temperature of the disinfectant is within the manufacturer's recommended limits.
9 Perform a high-level disinfection of the transducer. Use the following procedure:
a
Immerse the transducer in a high-level disinfectant solution. See the chemical manufacturer's
recommendations for soak time.
WARNING
Cautions
b
Ensure that the connector and most of the cable remain out of the fluid. You may submerge 2 inches
(5 cm) of the cable proximal to the transducer scanhead.
Connector
Do not submerge
130
Table 8-4, "High-level disinfectants compatible with
High-level disinfectants can cause harm to the patient if not completely removed
from the transducer. Follow the manufacturer's rinse instructions to remove
chemical residue.
Do not soak the transducer longer than recommended by the chemical
manufacturer.
Do not immerse the transducer connector in any disinfectant solution.
Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a
non- recommended disinfecting solution or incorrect solution strength can
damage or discolor the transducer and void the warranty.
Cable
Cleaning and disinfecting
Maximum
5 cm
(2 inches)
Scan head