Chapter 8: Cleaning and disinfecting the NanoMaxx
ultrasound system
Before getting started ................................................................................................................................ 126
Cleaning and disinfecting .......................................................................................................................... 127
Determining the required cleaning and disinfecting level ......................................................................................... 127
Spaulding classifications .......................................................................................................................................................... 127
Clean and disinfect system and transducer to a high level (semi-critical uses) ................................................... 128
Clean and disinfect system and transducer to a low level (non-critical uses) ...................................................... 131
Storing the transducer ............................................................................................................................... 133
Transporting the transducer ..................................................................................................................... 134
Compatible cleaners and disinfectants (system and transducers) ......................................................... 136
Compatible cleaners and disinfectants (system accessories) ................................................................. 139
Compatible high-level disinfectants (transducers) ................................................................................. 142
This chapter contains instructions for cleaning and disinfecting the NanoMaxx ultrasound system,
transducers, and accessories. Use the FUJIFILM SonoSite recommendations when cleaning or
disincomponentfecting your ultrasound system, transducer, and accessories. Use the cleaning
recommendations in the component manufacturer's instructions when cleaning or disinfecting your
components.
The system and transducers must be cleaned and disinfected after each exam. It is important to follow these
cleaning and disinfecting instructions without skipping any steps.
See
www.sonosite.com/transducers
SonoSite has tested and deemed compatible with the NanoMaxx system, transducers, and accessories.
This information is presented in the following tables.
Table 8-2, "Cleaners and disinfectants compatible with NanoMaxx system and
page 136 shows compatibility between disinfectants and the system.
Table 8-3, "Cleaners and disinfectants compatible with NanoMaxx system
shows compatibility between disinfectants and transducers.
Table 8-4, "High-level disinfectants compatible with NanoMaxx
compatibility between high-level disinfectants and transducers.
Note
Ensure that cleaning and disinfecting solutions and wipes are not expired
WARNING
Some cleaners and disinfectants can cause an allergic reaction to some individuals.
for images of transducers. Also included is a listing of disinfectants that
The system and transducers must be cleaned and disinfected after each exam. It is
important to follow these cleaning and disinfecting instructions without skipping
any steps.
transducers"on
accessories"on page 139
transducers"on page 142 shows