2 Immerse the transducer and about two inches of the cable in the disinfectant solution.
Caution
Use this diagram as a reference and then follow the instructions on the disinfectant label for the duration of
the transducer immersion.
3 Using the instructions on the disinfectant label, rinse to the point of the previous immersion, and then air
dry or towel dry with a clean, non-linting cloth.
4 Examine the transducer and cord for damage, such as cracks or splitting where fluid can enter.
If damage is evident, discontinue use of the transducer and contact FUJIFILM SonoSite or your local
representative
Table 2-4: Approved high-level disinfectants and soak times
Disinfectant
Cidex
Cidex OPA
Cleaning and disinfecting the transducer requires that you chose the proper cleaning and disinfecting level.
12
Do not immerse the transducer connector in any liquid.
Temperature
25° C (77° F)
20° C (68° F)
Duration
45 minutes
12 minutes
Determining the required cleaning and disinfection level