Level B: High-level disinfection (semi-critical)

Clean and disinfect the transducer immediately after use.
Be sure to follow the manufacturer's instructions when using cleaners and disinfectants.
Refer to
Table 2-3, "Compatible cleaners and
disinfectants recommended for use with the transducer.
Table 2-3: Compatible cleaners and disinfectants
Product
Sani-Cloth AF3 (grey top)
Sani-Cloth Bleach Wipes (yellow top)
Sani-Cloth Plus (red top)
Sani-Cloth HB (green top)
T-Spray II (PI-Spray II)
Cleaning and disinfecting the transducer
Use the following procedures to clean and disinfect the transducer.
To clean the transducer
1 Clean the transducer cord using either a cloth dampened with cleaner or disinfectant or a premoistened
disinfectant wipe. Start at the connector and wipe toward the scan head. Do not get any water or cleaner
on the connector.
2 Clean all transducer surfaces using either a soft, damp, soapy cloth or disinfectant wipe. Be sure to remove
any gels or particulate matter.
10
disinfectants"on page 10 for a list of cleaners and
P21v transducer
Determining the required cleaning and disinfection level