Working With The Product - Atmos MEDAP FINA VAC T 50 Operating Instructions Manual

Tapping unit for drainage
Table of Contents

Advertisement

4.2

Working with the product

Fig. 10: Setting the vacuum level
DANGER!
Infection hazard!
In the event of oversuction, the hydrophobic bacterial and viral filter must no longer
be used.
Replace the hydrophobic bacterial and viral filter with a new one.
WARNING!
Vacuum setting!
Make the vacuum settings very carefully! The regulating mechanism is sensitive.
Close the control valve very gently only.
WARNING!
Air inlet!
The drill hole on the underside of the unit must always be kept free so that airflow
is always ensured. The same applies to the four shunt air openings on the front of
the control valve.
NOTE
Check the vacuum setting once again immediately before using the unit!
3 3
2 2
Setting the vacuum for treatment
 Close the control valve (1) until it stops.
Turn in a clockwise direction.
 Bend the tube A (2) leading to the septic
fluid jar and set the vacuum with the control
valve while doing so. Turn anticlockwise.
Read the value on the vacuum gauge (3).
In case of malfunctions, e.g. insufficient
vacuum level, refer to the troubleshooting
table for the cause.
 Use the control valve to set the required
value for the treatment.
Increase the vacuum: Turn the control
1 1
valve anticlockwise.
Reduce the vacuum: Turn the control
valve clockwise.
Operation
Working with the product
GA 5752 3075 GB 13