Performance Tests
Pulse Modulation Rise Time Performance Test (Option 1E6 Only)

Measure the Rise Time

1. Use the timebase delay to align a rising pulse edge with the center of the oscilloscope
screen.
2. Change the timebase to 2 ns/div.
3. Display the envelope of channel 1.
4. Set the frequency of the 8648 to that listed in the test record. For this frequency
measure the 10% to 90% rise time of the RF pulse.

Test Record

• 8648B:
Table 8-43.
• 8648C:
Table 8-68.
• 8648D:
Table 8-93.