Service
Recommended Test Equipment
Table 5-1. Recommended Test Equipment
Instrument
Frequency
Resolution: 0.1 Hz
Counter
Function
Freq. range: 1 kHz
Generator
Amplitude: 4 Vpk
DC output: ±4 Vdc
Measuring
FM Accuracy: ±2% of reading ±1 digit
Receiver
AM accuracy: ±2% of reading ±1 digit
Range: 250 kHz to 1000 MHz
Filters: 300 Hz high-pass; 15 kHz
low-pass
Detectors: Peak+
Oscilloscope
Bandwidth: 1 GHz
Oscilloscope
Bandwidth: 100 MHz
Power Meter
Instrumentation accuracy: ±0.5%
Power reference accuracy: 0.9%
Power Sensor
Freq. range: 100 kHz to 4.2 GHz
Power range: −30 dBm to +13 dBm
Maximum SWR:
100 kHz to 300 kHz 1:1.6
300 kHz to 1 MHz 1:1.2
1 MHz to 2 GHz 1:1.1
2 GHz to 4.2 GHz 1:1.3
Cal factor accuracy (RSS): ≤1.6%
Power Sensor
Freq. range: 100 kHz to 4.2 GHz
Power range: −70 dBm to −20 dBm
(Low)
Maximum SWR:
100 kHz to 300 kHz 1:1.2
300 kHz to 2 GHz 1:1.15
2 GHz to 4.2 GHz 1:1.4
Power linearity (−30 to −20 dBm): ±1%
Cal factor accuracy: < 1.6%
Primary
Frequency: 10 MHz
Frequency
Stability: > 1 x 10
Standard
5-4
Critical Specifications
−10
/year
Recommended
Model
Performance Test - (P)
Supp. Verification Test - (V)
5316B
Internal Reference Oscillator
(Manual Adjustment)
33120A
AM Level and Distortion
AM Level
AM Level:FE
AM Modulator
Motherboard Audio Path
8902A
FM Accuracy
FM Distortion
AM Accuracy
Phase Modulation Distortion
Residual FM
RF Level Accuracy
AM Modulator
Motherboard Audio Path
54100A
Pulse Modulation On/Off Ratio
Pulse Modulation Rise Time
54600B
Internal Timebase: Aging Rate
438A
RF Level Accuracy
Power Level Accuracy
Detector Offset
Output Level
Predistortion& Detector Offset
Prelevel
Output Level:FE
HF Power Level Accuracy
8482A
RF Level Accuracy
Power Level Accuracy
Detector Offset
Output Level
Predistortion & Detector Offset
Prelevel
Output Level: FE
HF Power Level Accuracy
8481D
RF Level Accuracy
Option H70
5071A or
Internal Timebase: Aging Rate
5061A/B
Internal Reference Oscillator
(Manual Adjustment)
Time Base DAC
(Automated Adjustment)
Use
Adjustment - (A)
(A)
(A)
(A)
(A)
(A)
(A)
(P)
(P)
(P)
(P)
(P)
(P)
(A)
(A)
(P)
(P)
(P)
(P)
(P)
(A)
(A)
(A)
(A)
(A)
(A)
(P)
(P)
(A)
(A)
(A)
(A)
(A)
(A)
(P)
(P)
(A)
(A)

Hide quick links: