Performance Tests

AM Accuracy Performance Test (Option 1E2 Only)

AM Accuracy Performance Test (Option 1E2 Only)

Connect the Test Equipment

Figure 8-8. AM Accuracy Equipment Setup for Option 1E2

Configure the Measuring Receiver

1. Reset
2. AM mode
3. Peak+ detector
4. 300 Hz high-pass filter
5. 15 kHz low-pass filter

Configure the 8648

1. Turn AM on; press
.
AM MOD ON/OFF
2. Press
until the modulation generator (Option 1E2) sine waveform is activated.
INT 1 kHz
3. Set the rate; press
.
1 kHz
4. Turn the RF output on; press
.
RF ON/OFF

Measure Depths

1. Enter the amplitudes, frequencies and depths shown in the test record.
2. Record the test results and compare the results to the limits in the test record.