Adjustments
Agilent Service Support Software
4. When you have finished reading the information stored in the motherboard, select the
Cancel button to exit the program.
The Get Data button can be used if you are checking the data stored on additional
motherboards.
Storing Information in the A3 motherboard
To store information to the A3 motherboard, follow these steps;
1. Open the HP8648 Motherboard Utility:
• For MS Windows version 3.x;
a. Open the Program Manager window.
b. Open the Service Support program group.
c. Select the HP8648 Motherboard Utility icon.
• For MS Windows 95, 98, NT:
a. Select Start.
b. Select Agilent Service Software for PC's.
c. Select HP8648 Motherboard Utility.
2. When the User Information dialog box requesting a password is displayed, type: 8648.
Then select the OK button.
3. The HP8648 Motherboard Repair Utility window is displayed so that the instrument
information can be edited and stored in the motherboard. Refer to Figure ?. Edit the
information in this window to reflect the actual changes.