Spurious Performance Test; Connect The Test Equipment; Configure The Spectrum Analyzer; Configure The 8648 - Agilent Technologies 8648A Operation And Service Manual

Signal generator
Hide thumbs Also See for 8648A:
Table of Contents

Advertisement

Spurious Performance Test

Connect the Test Equipment

Figure 8-14. Spurious Equipment Setup

Configure the Spectrum Analyzer

1. Frequency span 500 kHz
2. Resolution BW 1 kHz
3. Video BW 1 kHz

Configure the 8648

1. Turn the RF output on; press
2. Turn modulation off; press

Measure Spurious Levels

1. Measure the levels of the spurious signals shown in the test record.
2. Convert the measured levels to decibels below the fundamental (dBc) and compare the
results to the corresponding limits.

Test Record

• 8648A:
Table 8-14.
• 8648B:
Table 8-35.
• 8648C:
Table 8-60.
• 8648D:
Table 8-85.
.
RF ON/OFF
.
MOD ON/OFF
Performance Tests
Spurious Performance Test