Supplemental Verification Tests

9 kHz RF Level Accuracy Supplemental Verification Test

9 kHz RF Level Accuracy Supplemental Verification Test

Connect the Test Equipment

Figure 9-2. 9 kHz RF Level Accuracy Equipment Setup

Configure the Multimeter (DMM)

• Function ACV
• Range AUTO
• Connect the Positive and Ground leads from the Input (2 Wire) connector across a 50Ω
resistor (1/4 W, 0.1% tolerance).

Configure the 8648

• Set the frequency; press
• Set the amplitude; press
• Connect the BNC cable to the RF OUTPUT of the 8648.
• Connect the BNC splitter to the other end of the BNC cable attaching:
— the splitter's positive lead to the side of the resistor that is connected to the HI lead
from the DMM
— the splitter's ground lead to the side of the resistor that is connected to the LO lead
from the DMM
• Turn on the RF output; press

Measure RF levels

1. Set the amplitudes shown in the test record.
9-8
,
,
.
FREQUENCY
9
kHz
,
,
AMPLITUDE
10
dBm
.
RF ON/OFF
.