Agilent Technologies 8648A Operation And Service Manual page 436

Signal generator
Hide thumbs Also See for 8648A:
Table of Contents

Advertisement

Performance Tests
8648C Test Record
NOTE
If the automated Power Level Accuracy performance test is performed, the
RF Level Accuracy performance test is not required.
Table 8-63. RF Level Accuracy Performance Test - Part 2
Frequency
Amplitude
(MHz)
(dBm)
2000
13
10
4
−5.9
−15.9
2499
13
10
4
−5.9
−15.9
3200
13
10
4
−5.9
−15.9
If the automated Power Level Accuracy performance test is performed, the
NOTE
8-118
Limits (dBm)
Lower
Measured
12
9
3
−6.9
−16.9
12
9
3
−6.9
−16.9
11.5
8.5
2.5
−7.4
−17.4
Uncertainty
Upper
(±dB)
14
0.15
11
0.15
5
0.15
−4.9
0.15
−14.9
0.12
14
0.32
11
0.32
5
0.32
−4
0.32
−14.9
0.25
14.5
0.46
11.5
0.46
5.5
0.46
−4.4
0.46
−14.4
0.41