Agilent Technologies 8648A Operation And Service Manual page 96

Signal generator
Hide thumbs Also See for 8648A:
Table of Contents

Advertisement

Operation Reference
Function
Vernier Ranges The following table, "10 dB Specified Vernier Ranges", provides the
upper and lower limits of each vernier range. The instrument's amplitude setting when
you press the
ATTN HOLD
The vernier is allowed to over-range and under-range beyond the limits shown in the table
when
is selected. However, amplitude settings that exceed the limits may not
ATTN HOLD
provide output levels that are within the accuracy specifications of the instrument.
3. RF OUTPUT
The following table shows the reverse power protection for the RF output port of each
instrument model. When the instrument senses a reverse-power signal, it turns the RF
output off, the step attenuator to maximum attenuation, and the vernier to its lowest
setting. A message appears in the second line of the display when the reverse power
protection has been activated.
After you have removed the reverse-power signal from the RF output, press the
key to turn the output signal on again.
1b-48
key determines which vernier range is used.
10 dB Specified Vernier Ranges
Range
Upper and Lower Limits
1
+10.0 dBm
−6.0 dBm to −15.9 dBm
2
−16.0 dBm to −25.9 dBm
3
−26.0 dBm to −35.9 dBm
4
−36.0 dBm to −45.9 dBm
5
−46.0 dBm to −55.9 dBm
6
−56.0 dBm to −65.9 dBm
7
−66.0 dBm to −75.9 dBm
8
−76.0 dBm to −85.9 dBm
9
−86.0 dBm to −95.9 dBm
10
−96.0 dBm to −105.9 dBm
11
−106.0 dBm to −115.9 dBm
12
−116.0 dBm to −127.0 dBm
13
−127.1 dBm to −136.0 dBm
14
a. +13 dBm for 8648B/C/D
a
to −5.9 dBm
RF ON/OFF