Agilent Technologies 8648A Operation And Service Manual page 57

Signal generator
Hide thumbs Also See for 8648A:
Table of Contents

Advertisement

FLEX/FLEX-TD
Setting the data Rate and Pager Type
The parameter menu for the data rate and the pager type looks like the following:
• DATA RATE specifies how fast the data stream is output in bits per second and whether it is
2-level or 4-level FSK.
1600/2:
1600 bps, 2-level FSK
3200/2:
3200 bps, 2-level FSK
3200/4:
3200 bps, 4-level FSK
6400/4:
6400 bps, 4-level FSK
• PAGER TYPE specifies the type of message a pager under test can respond to and is selectable
from the following:
Short Message/Tone only type
TONE ONLY:
Numeric type
NUMERIC:
ALPHANUMERIC: Alphanumeric type
HEX/Binary type
HEX/BIN:
If NUMERIC is selected, the next setting to be selected is VECTOR TYPE as shown:
• VECTOR TYPE identifies the type of message as follows:
The pager-under-test displays the numeric message it receives.
STANDARD:
The pager-under-test converts the numeric message to a pre-defined message
SPECIAL:
which is displayed.
The message numbers are assigned for each paging address separately
NUMBERED:
starting at 0 and progressing up to a maximum of 63 in consecutive order.