Am Accuracy Performance Test; Connect The Test Equipment; Configure The Measuring Receiver; Configure The 8648 - Agilent Technologies 8648A Operation And Service Manual

Signal generator
Hide thumbs Also See for 8648A:
Table of Contents

Advertisement

Performance Tests

AM Accuracy Performance Test

AM Accuracy Performance Test

Connect the Test Equipment

Figure 8-7. AM Accuracy Equipment Setup

Configure the Measuring Receiver

1. Reset
2. AM mode
3. Peak+ detector
4. 300 Hz high-pass filter
5. 15 kHz low-pass filter

Configure the 8648

1. Turn AM on; press
2. Set the rate; press
3. Turn the RF output on; press

Measure Depths

1. Enter the amplitudes, frequencies and depths shown in the test record.
2. Record the test results and compare the results to the limits in the test record.

Test Record

• 8648A:
Table 8-8.
• 8648B:
Table 8-28.
• 8648C:
Table 8-53.
• 8648D:
Table 8-78.
8-16
.
AM MOD ON/OFF
.
INT 1 kHz
RF ON/OFF
.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8648b8648c8648d3847a3847u4037a ... Show all

Table of Contents