Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
4.2.2

Manifold Light

Clean the Ettan IPGphor Manifold Light with water, no warmer than 40 °C, and
with IPGphor Strip Holder Cleaning Solution.
CAUTION! Do not use any organic solvents.
4.2.3

Electrodes

Wash the electrode assemblies with the IPGphor Strip Holder Cleaning Solution.
Rinse thoroughly with water then deionized water and allow to air dry.
CAUTION! Do not use strong acids, bases, ketones, alcohols, or other reagents
to clean the covers, sample cups, or electrodes or the parts may be damaged.
Cups may be briefly rinsed with ethanol if desired.
Care and maintenance
4
51