Instrument set up

Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
2.1

Placing the instrument

WARNING! The vents at the front, back, and bottom of the instrument must
not be obstructed. Air must be free to circulate for the cooling apparatus to
function properly. Minimum free space behind the rear panel should be
10 cm.
IMPORTANT! Select a place for Ettan IPGphor 3 where it is possible to access
the mains power switch on the rear panel of the instrument.
Position the instrument on a flat surface with the spirit level in the center. Turn the
leveling feet as necessary to make the instrument level and stable.
2.2

Power connection

Two power cords are supplied: one cord has a "North American" style (UL817)
3-pronged plug for 115 V~ power outlets, and the other has a "Central European"
style (CEE7/VII) plug for 230 V~ power outlets.
1
Select the cord with the appropriate plug for the power outlet in your
laboratory. If neither of these cords are suitable for your power outlets,
obtain a detachable cord with an IEC/320/C13 (CEE22/V) outlet.
2
Plug the connectors into the power cord receptacle on the rear panel and
into a properly grounded power outlet.