X-STREAM X2FD
2.2

Installing the Analyzer

Installing and wiring this instrument must comply with all relevant national
legislative requirements and regulations.
Consider all safety instructions within this on hand manual and all associa-
ted analyzer instruction manuals!
Installing this instrument requires opening the enclosure and working at
the open instrument. This is permitted only when both no hazardous atmos-
phere is present and the instrument and connected external circuitry are
de-energized!
Depending on the local regulation, this may require a competent hot work
supervisor to issue a hot work permit.
The analyzer model X-STREAM X2FD, to which this manual relates, intended
to be wall mounted and/or outdoor installed, weighs up to approx. 63 kg
(139 lbs), depending on included options!
Use two people and/or suitable tools for transportation and lifting these
instruments!
Take care to use anchors and bolts specifi ed to be used for the weight of
the units!
Take care the wall or stand the unit is intended to be installed at is solid and
stable to hold the units!
2-2
2
Installation
POSSIBLE EXPLOSION HAZARD
POSSIBLE EXPLOSION HAZARD
HEAVY INSTRUMENT
Instruction Manual
HASXMDE-IM-EX
Emerson Process Management GmbH & Co. OHG
01/2015