1.5 Explosion Protection Compliances - Emerson ROSEMOUNT Xstream X2 Series Instruction Manual

Gas analyzers
Hide thumbs Also See for ROSEMOUNT Xstream X2 Series:
Table of Contents

Advertisement

X-STREAM X2FD

1.5 Explosion Protection Compliances

This product is available in two different varia-
tions, separately certifi ed by agencies for the
use in hazardous (classifi ed) areas:
The one version, to be equipped with cable
glands, is certifi ed by Fyzikálně technický
zkušební ústav, s.p (FTZÚ), an European
Notifi ed Body under the Directive 94/9/EC
(„ATEX") and conforms to the provisions of
EN 60079-0 and EN 60079-1. See appen-
dix for a copy of the EC type examination
certifi cate.
The second variation, to be equipped with
metric-to-NPT adapters and conduits (these
components are not part of the instrument
certifi cation), is certifi ed by the Canadian
Standards Association, an „OSHA Na-
tionally Recognized Testing Laboratory"
(NRTL), for Canada and USA and conforms
to the provisions of CAN/CSA-E60079-0:02
(R2006), CAN/CSA-E60079-1:02 (R2006),
ANSI/ISA-12.00.01-2002 (IEC 60079-0 Mod),
ANSI/ISA-12.22.01-2002 (IEC 60079-1 Mod).
Furthermore, these X-STREAM X2FD analy-
zers are certifi ed for use in Class I, Division 2,
Group BCD T3 areas.
See appendix for a copy of the CSA Certifi cate
of Compliance.
IECEx certifi cation enables worldwide appro-
vals with minimized need of testing. Further-
more, based on this approval several national
approvals are granted.
For a comprehensive list of applicable certi-
fi cates visit our website at
www.emersonprocess.com.
1-4
1.5 Compliances
The following certifi cation markings apply to
the products:
European Union (EU, ATEX)
Category 2, Zone 1:
Ex d IIB+H
EC ATEX Type Examination Certifi cate:
FTZU 08 ATEX 0028 X.
IECEx
Ex d IIB+H
Conforms to the provisions of the „Equipment
intended for use in Potentially Explosive At-
mospheres (ATEX)" Directive 94/9/EC, EMC
Directive 2004/108/EC and CE Directive
93/68/EEC.
USA
Class I, Zone 1, AEx d IIB+H
Class I, Division 2, Group BCD T3
Canada
Class I, Zone 1, Ex d IIB+H
Certifi cate of Compliance 1714037X
Instruction Manual
HASXMDE-IM-EX
T4 Gb
2
T4 Gb
2
2
T3
2
Emerson Process Management GmbH & Co. OHG
01/2015
T3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount xstream x2fd