INFICON
9.1.4 Quick Reference SOP- Heat-up and Tune
9.2 Select a Different Method Using the SELECT
074-471-P1E HAPSITE ER Operating Manual
CAUTION
Do not open the front panel in a wet or contaminated area.
1
Insert the internal standard and carrier gas canisters.
2
Insert a charged battery.
3
Connect the AC to DC power converter power supply.
4
Verify that the appropriate sample configuration (i.e., concentrator) is installed.
5
Press the POWER button on the front panel.
6
HAPSITE ER will heat the necessary components and perform AutoTune. A
prompt to run SURVEY or ANALYZE will appear when HAPSITE ER is ready
to run a sample.
7
If the default method is not the desired method, touch STOP PREPARE.
8
Touch SELECT METHOD. Highlight the desired method. Touch SELECT.
When the SYSTEM READY message is displayed, touch either RUN SURVEY or
RUN ANALYZE. If using the push buttons, push SURVEY RUN or ANALYZE RUN.
If connecting wirelessly to the laptop, see Communications and Touch Screen Options
[} 110].
METHOD Icon
If the default method is not the desired method, the method can be changed.
Changing the method can occur when the system is preparing or when another
method has finished preparing.
1
When the PREPARING SYSTEM screen is displayed, touch STOP PREPARE.
Alternately, use the arrow keys to highlight STOP PREPARE and push OK
SEL.
Operate the Instrument in Portable Mode | 9
71 / 436
Need help?
Do you have a question about the HAPSITE ER and is the answer not in the manual?