Fluke 6135A/PMU Operator's Manual
Fluke 6135A/PMU Operator's Manual

Fluke 6135A/PMU Operator's Manual

Calibration system
Table of Contents

Advertisement

Quick Links

6135A/PMU
Calibration System
Operators Manual
October 2013 Rev. 1, 2/2020
© 2013 – 2020 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.

Advertisement

Table of Contents
loading

Summary of Contents for Fluke 6135A/PMU

  • Page 1 6135A/PMU Calibration System Operators Manual October 2013 Rev. 1, 2/2020 © 2013 – 2020 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
  • Page 2 Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
  • Page 3: Table Of Contents

    Server PC ....................... 1-11 6135A Three Phase Power Standard ............. 1-12 Microsemi GPS Receiver ................1-14 6135A/PMU System Timing Unit ..............1-15 6135A/PMU System Software Status Panel ............1-17 PMUCal Software ....................1-18 Overview ....................... 1-18 Monitor Screen ....................1-18 Navigation Menu ...................
  • Page 4 6135A/PMU Operators Manual Input Step Change Tests ................1-22 Installation Instructions ..............2-1 About this Chapter ..................... 2-3 Unpack and Inspect .................... 2-3 Install the Calibration System ................2-4 Pre-Installation Preparation ................2-4 Installation Procedure ..................2-5 Install the PMUCal Software ................2-5 Configure the Client PC ..................
  • Page 5 Contents (continued) Test Report and Calibration Report ..............3-27 UUT Calibration Setup ................ 4-1 About this Chapter ..................... 4-3 Safety Information ..................... 4-3 General Operation Information ................4-4 Turn On the Calibration System ..............4-4 Turn Off the Calibration System ..............4-4 Calibration System Warm-up ................
  • Page 6 Troubleshooting, Maintenance, and Care ......... 8-1 About this Chapter ..................... 8-3 Troubleshooting ....................8-4 General ......................8-4 6135A/PMU System Timing Unit ..............8-4 Server PC ....................... 8-5 6135A Three Phase Power Standard ............. 8-5 GPS Receiver ....................8-6 Network ......................8-6 Calibration Software ..................
  • Page 7: Introduction And Specifications

    (see Figure 1-1). See the “Components of the Calibration System” for more information about each individual device. The Fluke 6105A and two 6106A together comprise a 6135A. The Fluke 6135A is a source of precision three phase voltage and current signals. These precise signals must be related to Universal Coordinated Time (UTC) to be useful for calibration of PMU UUTs.
  • Page 8 Server PC (the “Server”) 6105A Electrical Power Standard Master Unit (the “6105A Master Unit”) 6106A Electrical Power Standard #1 (the “6106A Auxiliary Unit 1”) 6106A Electrical Power Standard #2 (the “6106A Auxiliary Unit 2”) gtu009.emf Figure 1-1. The 6135A/PMU Calibration System...
  • Page 9: Standard Accessories

    The Calibration System comes with items in Table 1-1. The accessories and manuals can be found in the boxes inside the shipment crate. To order replacement parts or accessories, contact Fluke Calibration and reference the part numbers and item names in Table 1-1.
  • Page 10: Safety Information

    About the Manual Set The Calibration System comes with an Operators Manual and Safety Information. These manuals are supplied on the CD and can be downloaded from Fluke Calibration’s website at http://us.flukecal.com/literature/manuals/search. This Operators Manual supplies all the information to safely unpack, install, configure, and use the Calibration System.
  • Page 11: Contact Fluke Calibration

    6135A Three Phase Power Standard to Fluke for calibration and verify the rest of the system with a few simple time measurements using an oscilloscope as instructed in Chapter 9. To schedule and send the...
  • Page 12: Calibration System Overview

    CONNECTED RUNNING TCP / IP COMMS RUNNING SERVICE RUNNING SERVICE RUNNING Input/Output Panel (Shown Above) 6135A/PMU System Timing Unit (the “Timing Unit”) Microsemi SyncServer S650 (the “GPS Reciever”) Server PC (the “Server”) 6105A Electrical Power Standard Master Unit (the “6105A Master Unit”)
  • Page 13 Introduction and Specifications Calibration System Overview GNSS CONSOLE (RS232) J3 (IRIG) J5 (10 MHz) gtu179.emf Figure 1-3. The Calibration System Connections...
  • Page 14: Components Of The Calibration System

    6135A/PMU Operators Manual Components of the Calibration System Client PC The Client PC is the personal computer used to operate the Calibration System. Note The Client PC is not supplied with the Calibration System. The Client PC must meet the minimum requirements listed in Table 1-4 and have the Calibration Software installed (see “Install the Calibration Software”...
  • Page 15: Server Pc

    Server PC The Server PC (shown in Figure 1-5) is a server computer that receives commands from the Client PC and controls the Calibration System. The default settings made by Fluke at the time of shipping are: PC name: PMUCal •...
  • Page 16: 6135A Three Phase Power Standard

    The 6135A System contains one 6105A Electrical Power Standard (L1) and two 6106A Electrical Power Standards (L2 and L3). The complete 6135A System is connected to and controlled by the 6135A/PMU System Timing Unit (the Timing Unit) and the Server PC (see Figure 1-3).
  • Page 17 Calibration System to a UUT” in Chapter 4. The Fluke 6135A/PMU Calibration System can support Fluke 6105A/6106As with 50 A or 80 A options or up to three Fluke 52120A Transconductance Amplifiers per channel. The higher current outputs are taken directly from the units outputs.
  • Page 18: Microsemi Gps Receiver

    6135A/PMU Operators Manual Table 1-6. 6135A Connections gtu010.emf Item Name Description  Voltage HI Output Positive voltage output terminal that supplies voltage to the UUT. Negative voltage output terminal that supplies voltage to the Voltage LO Output  UUT. Current HI Output Positive current output terminal that supplies current to the UUT.
  • Page 19: 6135A/Pmu System Timing Unit

    System. 6135A/PMU System Timing Unit The 6135A/PMU System Timing Unit (the Timing Unit) is the timing and modulation control device in the Calibration System. This device is connected to the Server PC, the 6135A System, the GPS Receiver, and the UUT (see Figure 1-3).
  • Page 20 6135A/PMU Operators Manual Table 1-8. The Timing Unit Connections gtu027.emf Item Name Description GPS Antenna Out GPS Antenna signal output to the UUT.  1PPS signal output to the UUT. The signal is set to pulse on the 1PPS Out ...
  • Page 21: 6135A/Pmu System Software Status Panel

    6135A/PMU System Software Status Panel 6135A/PMU System Software Status Panel The 6135A/PMU System Software Status Panel (the Status Panel) is located on the system cabinet door and indicates which software services are active. Table 1-10 shows the Status Panel and describes the indicators.
  • Page 22: Pmucal Software

    6135A/PMU Operators Manual PMUCal Software Overview The PMUCal Software is the control interface to the Calibration System to test and calibrate a UUT. The PMUCal Software is a program that is installed the Client PC. Read the subsequent sections to learn about the basic menus and indicators of the PMUCal Software.
  • Page 23: Navigation Menu

    Introduction and Specifications PMUCal Software Navigation Menu The navigation menu supplies you with all the buttons necessary to navigate the Calibration Software. See Table 1-11 for detailed information on each button in the navigation menu. Table 1-11. The Navigation Menu Button Name Description Navigates to the test menu screen to run a test.
  • Page 24: The System Status Panel

    Illuminates when the Client PC connects to the 6135A System through the Server PC. Flashes when the 6105 is being configured for a test run. Illuminates when the Client PC connects to the 6135A/PMU System Timing Unit through the Server PC.
  • Page 25: General Specifications

    Introduction and Specifications General Specifications General Specifications Input Power Voltage ..............100 V to 240 V with up to ±10 % fluctuations Frequency .............. 47 Hz to 63 Hz Maximum consumption ........3,500 VA Max Dimensions Size Height..............1,170 mm (46 in) Width ..............
  • Page 26: Pmu Test: Limits To Range Of Influence Quantities

    6135A/PMU Operators Manual PMU Test: Limits to Range of Influence Quantities General Test limits Influence Quantity Range (voltage) Range (current) PMU Nominal Frequency 50 Hz or 60 Hz 50 Hz or 60 Hz PMU Nominal Magnitude 10 V to 240 V...
  • Page 27: Installation Instructions

    Chapter 2 Installation Instructions Title Page About this Chapter ..................... 2-3 Unpack and Inspect .................... 2-3 Install the Calibration System ................2-4 Pre-Installation Preparation ................2-4 Installation Procedure ..................2-5 Install the PMUCal Software ................2-5 Configure the Client PC ..................2-6 Configure Windows Firewall ................
  • Page 28 6135A/PMU Operators Manual...
  • Page 29: About This Chapter

    Unpack and Inspect The Calibration System is shipped in a wood shipment crate. Use the shipment crate to transport or ship the calibrator to Fluke Calibration for adjustment or repair. Note After the Calibration System is removed from the crate, store the crate in a dry, safe area for future use.
  • Page 30: Install The Calibration System

    6135A/PMU Operators Manual Install the Calibration System This section supplies information on how to install the Calibration System. Use this section for the initial installation and also when the Calibration System comes back from annual adjustment or repair. Pre-Installation Preparation Placement preparation is important for a successful Calibration System installation.
  • Page 31: Installation Procedure

    Installation Instructions Install the PMUCal Software Installation Procedure After the Calibration System has been unpacked and the pre-installation is complete, install the Calibration System as follows: 1. Carefully move the Calibration System to the installation location. 2. Lower the system feet to prevent Product movement. 3.
  • Page 32: Configure The Client Pc

    6135A/PMU Operators Manual gtu044.jpg Figure 2-1. Installation Destination Directory Screen 1. Read the License Agreements and click Accept or Decline . 2. Click Next >> to start the installation. 3. Click Finish to exit the installation software. 4. Click Restart to restart your computer.
  • Page 33 Installation Instructions Configure the Client PC gtu162.jpg 2. Click Allow another program…. gtu163.jpg 3. Select PMUCalClient from the list. gtu164.jpg 4. Click Add .
  • Page 34 6135A/PMU Operators Manual 5. Scroll down to PMUCalClient and make sure all check boxes are checked. gtu165.jpg Note The Domain column only shows if the PC is connected to a network that is part of a domain. 6. Click OK .
  • Page 35: Network Connection Instructions

    Installation Instructions Configure the Client PC Network Connection Instructions The Client PC and the Calibration System can be connected to a network. If a network is not available, use the instructions in the “Direct Connection Instructions” section to make a PC to PC connection. Connect the Calibration System to the network using a standard Category 5 (CAT-5) straight-through Ethernet cable to the Input/Output Panel Ethernet connector (above GPS antenna connection).
  • Page 36: Direct Connection Instructions

    6135A/PMU Operators Manual Direct Connection Instructions When a network is not available, the Calibration System can be connected directly to a To set up a direct connection: 1. Connect the Client PC directly to the Server PC. To do this, connect one end of an Ethernet cable to right most Ethernet port on the Input/Output Panel.
  • Page 37 Installation Instructions Configure the Client PC d. On the Network Adapter menu, right-click on the Local Adapter and then select Properties from the list. gtu169.jpg e. On the Local Area Connection Properties menu, highlight Internet Protocol Version 4 (TCP/IPv4) then click the Properties button. gtu170.jpg Select the Alternate Configuration tab.
  • Page 38 6135A/PMU Operators Manual Note The Alternate Configuration tab only shows if the PC is set up to automatically obtain an IP address (DHCP). If the PC has an IP assigned by the network administrator, use the address in the General tab. Record the current addresses so they will not be lost.
  • Page 39 Installation Instructions Configure the Client PC 4. Change the location of the network as follows: Note Skip this step if the location of the network was previously set. a. Open the Windows Start Menu and type “network and sharing” in the search box.
  • Page 40: Set Up The File Storage Directory

    Settings screen will be displayed. Refer to Table 2-1. Note Fluke recommends that you use the default folder locations. If a different folder is necessary, click on the folder icon to the right of the Working Directory Path and choose a new folder. A Windows Explorer dialog box opens.
  • Page 41 Installation Instructions Configure the Client PC Table 2-1. Folder Location Settings tu135.emf Item Name Description Path of the top-level directory that contains the settings, tests, Working (top level) folder  results, and reports subdirectories. Path of the directory that contains the compliance tests made Test Subfolder with the test maker tool (refer to “About the Test Files and Test ...
  • Page 42: Install The Test Files

    6135A/PMU Operators Manual Install the Test Files The test files used by the system are supplied separately. They can be found in a .zip file in the Test Files folder on the Calibration System software installation CD-ROM. To install the test files: 1.
  • Page 43: Software Operation

    Chapter 3 Software Operation Title Page About this Chapter ..................... 3-3 UUT Calibration Procedure ................3-3 Overview ....................... 3-3 Calibration Procedure Information ..............3-4 Calibration Compliance to IEEE C37.118.1a-2014 ........3-5 About Synchrophasor Calibration ..............3-6 Time to Complete a UUT Calibration ............3-7 Calibration Test Measurements ..............
  • Page 44 6135A/PMU Operators Manual...
  • Page 45: About This Chapter

    Software Operation About this Chapter About this Chapter This chapter supplies information on the PMUCal Software supplied with the Calibration System. The beginning of the chapter provides information on the UUT calibration procedure that is controlled by the PMUCal Software. This information is followed by functionality and feature information to help understand the different testing methods and modes of operation available.
  • Page 46: Calibration Procedure Information

    6135A/PMU Operators Manual Calibration Procedure Information The UUT calibration procedure illustrated in Figure 3-2 is a remotely operated procedure that is controlled by the PMUCal Software. To certify a UUT, the PMUCal Software is configured to run preprogrammed automated certification tests (the certification tests). A certification test is an automated test file that contains a noneditable list of sequenced tests (refer to “Automated Testing”...
  • Page 47: Calibration Compliance To Ieee C37.118.1A-2014

    UUT Calibration Procedure Calibration Compliance to IEEE C37.118.1a-2014 The Fluke 6135A PMU Calibration System was developed to provide PMU manufacturers and third party calibration organizations with a system to demonstrate compliance to IEEE C37.118.1a-2014 (the Standard). The Calibration System is more than 10 times more accurate than the PMU test limits in the Standard.
  • Page 48: About Synchrophasor Calibration

    6135A/PMU Operators Manual About Synchrophasor Calibration For a PMU to generate synchrophasors, a PMU must either have its own built in time reference or take Coordinated Universal Time (UTC) from an external device. UTC is most commonly obtained from a GPS device and the PMU may receive satellite data from an aerial.
  • Page 49: Time To Complete A Uut Calibration

    UUT Calibration Procedure “True” STU_T0 and STU_TRefp Synchophasor Mathematical Model Hardware Trigger Aerial Nominal IRIG-B 30 MHz System Results = Clock Fluke Timing Unit Computed 6135 A 10 MHz (STU) PMU Errors Analogue Signals Receiver 1 pps Aux(T IRIG-B Measured Synchophasor gtu177.emf...
  • Page 50: Calibration Procedure Preparation

    6135A/PMU Operators Manual Calibration Procedure Preparation Before the calibration procedure is started, it is necessary to configure the UUT and also set up the UUT personality in the PMUCal Software. The information in this section describes the differences between the UUT configuration and the UUT personality profile.
  • Page 51 Software Operation Calibration Procedure Preparation gtu032.png Figure 3-6. UUT Personality Screen (Information Tab) gtu033.png Figure 3-7. UUT Personality Screen (Phase Order)
  • Page 52: Simulator Personality Profile

    6135A/PMU Operators Manual Simulator Personality Profile The Calibration Software has a default simulator personality profile (the simulator profile) to change the personality of the built-in UUT simulator. The simulator profile loads each time you choose Simulation Mode on the main startup window. Unlike the UUT profile, the simulator profile lets the operator change the UUT configuration items such as nominal frequency, reporting rate, class, nominal voltage, and nominal current.
  • Page 53: Testing Methods

    Software Operation Testing Methods Testing Methods A UUT can be tested with one of three methods: automated testing, general interactive testing, or custom interactive testing. To certify and verify a UUTs performance, the automated testing method must be used. Selection of a testing method depends on the test requirements and what format test results must be in.
  • Page 54 6135A/PMU Operators Manual Figure 3-9 illustrates the automated test and interactive test process. For instructions on how to do an automated test or a general test, see Chapter 5 and Chapter 6. See the subsequent sections for more information on each testing method.
  • Page 55: Automated Testing

    Software Operation Testing Methods Automated Testing An automated test runs a list of tests one-by-one without the need for operator interaction. As the tests are completed, the Calibration Software saves or overwrites the maximum test results values to the active results file (refer to “Standard Compliance Tests and Test Files”...
  • Page 56: General Interactive Testing

    6135A/PMU Operators Manual General Interactive Testing General interactive testing (see Figure 3-11) is a basic test method that loads a single, preconfigured test file and interactively graphs the test results. When Run Single Test is clicked, the PMUCal Software loads a test from a file that you select and runs the test with the test parameters saved in the file.
  • Page 57: Interactive Test Sequence Of Events

    Software Operation Testing Methods gtu072.png Figure 3-12. Test Configuration Window Interactive Test Sequence of Events Run the Test When Customize and Run Single Test is clicked, the PMUCal Software opens the test configuration window. From here, the test parameters can be changed and the test can be run or simulated.
  • Page 58: Set Up The Interactive Graph

    6135A/PMU Operators Manual Set Up the Interactive Graph After the test is complete, the graph configuration window (see Figure 3-14) opens that lets the operator configure the test results in the interactive graph (see “Interactive Graphs, Results, and Reports” in Chapter 8). When Plot Results is clicked, the PMUCal Software reads the test results data and makes the interactive graph.
  • Page 59: Modes Of Operation

    Software Operation Modes of Operation measurement values (set the graph dimensions, zoom-in and zoom-out both vertically and horizontally, change the plot types, and adjust the colors of the plot). For instructions on how to use these functions, see Chapter 7. gtu075.jpg Figure 3-15.
  • Page 60: Simulation Mode

    6135A/PMU Operators Manual Simulation Mode The Calibration Software comes with a fully programmed virtual PMU referred to as the PMU simulator. The PMU simulator has a default named personality profile (see “Simulator Personality Profile” in this Chapter).The PMU simulator is programmed to comply with IEEE C37.118.1a-2014, Annex C, Reference Signal Processing Models.
  • Page 61 Software Operation Standard Compliance Tests and Test Files Main PMUCal Top-Level Folder Tests Folder Automation 50 Hz or 60 Hz (UUT nominal frequency) Latency (compliance test) Frequency Ramp (compliance test) Step Tests Modulation (compliance test) Latency Amplitude Combined (compliance subtests) Phase Steady State (compliance test) Harmonic...
  • Page 62: Standard Compliance Tests And Subtests

    6135A/PMU Operators Manual Standard Compliance Tests and Subtests IEEE C37.118.1a-2014, Synchrophasor Measurements for Power Systems Specification is the standard that mandates which tests are required to certify and verify UUT performance. The three compliance tests types are steady state, dynamic, and message.
  • Page 63: Dynamic Compliance Tests

    Software Operation Standard Compliance Tests and Test Files Dynamic Compliance Tests A dynamic compliance test type is where the magnitude, frequency, and phase offset vary depending on the type of test. Dynamic compliance tests make sure that the UUTs performance is within the range and limitations in IEEE C37.118.1a-2014 “Dynamic Compliance”.
  • Page 64: Standard Compliance Tests File Naming Convention

    6135A/PMU Operators Manual Standard Compliance Tests File Naming Convention The standard compliance tests are saved in a folder named with the compliance test and subtest name (for example, “..\Tests\60Hz\SteadyState\Magnitude\”). Inside the folder, each compliance test is named with the primary test parameters in the filename to help visually identify the test parameters for the test.
  • Page 65 Software Operation Standard Compliance Tests and Test Files gtu064.jpg Figure 3-19. Dynamic Compliance Subtest Naming Convention Latency Sub-tests Naming Convention Measurment Latency Sub-test: File Path: ..\ Tests\<50Hz or 60Hz>\Latency Example: 60f0_30fs_m.xml Reporting Rate 00f0_00fs_X_.xml Nominal PMU Class Frequency (M or P) gtu144.emf Figure 3-20.
  • Page 66: Automated Test List Files

    6135A/PMU Operators Manual Automated Test List Files The automated test list files are saved in “Automated” folder. Figure 3-21 illustrates the automated test list naming convention. AutomatedTest List Files Naming Convention Automated Test List: File Path: ..\ Tests\Automation Example: 50F0_10Fs_P_Complete.xml...
  • Page 67: Results Files Naming Convention

    Software Operation Test Results and Results Files The active test results file values can be seen on the results screen (see Figure 3-23). The maximum values are used to verify that the performance parameters are in the specified range limits. If test data was not recorded for a test or subtest, then data is not shown for that test.
  • Page 68: Result File Save Options

    After the initial installation of the PMUCal Software, it is necessary to choose where to make the top-level folder named Fluke PMUCal. Once made, the folder will contain a folder named “results”. The results folder is the storage location for all results files.
  • Page 69: Raw Data Files

    Software Operation Test Report and Calibration Report Main PMUCal Top-Level Folder Results Folder UUT Personality Results Folder Raw Data gtu060.emf Figure 3-26. Test Results Folder Structure Raw Data Files Raw test data is the full set of unedited and unformatted test data files in Comma Separated Values (CSV) format that the system collected from the test.
  • Page 70 6135A/PMU Operators Manual 3-28...
  • Page 71: Uut Calibration Setup

    Chapter 4 UUT Calibration Setup Title Page About this Chapter ..................... 4-3 Safety Information ..................... 4-3 General Operation Information ................4-4 Turn On the Calibration System ..............4-4 Turn Off the Calibration System ..............4-4 Calibration System Warm-up ................ 4-4 Connect the Calibration System to a UUT ............
  • Page 72 6135A/PMU Operators Manual...
  • Page 73: About This Chapter

    UUT Calibration Setup About this Chapter About this Chapter This chapter supplies information necessary to safely connect the Calibration System to a UUT. Each UUT has unique features that will affect this connection and configuration procedure. Before this procedure is started, read the UUT user documentation to fully understand the operation, limits, and features of the UUT.
  • Page 74: General Operation Information

    2. Push the Power button on the rear panel of the GPS Receiver (if required). 3. Push the Power button on the 6135A/PMU System Timing Unit front panel. 4. Push the Power button on the two 6106A (Auxiliary) L2 and L3 units . Wait a few minutes then push the Power button on the 6105A (Master) L1 unit .
  • Page 75: Prepare The Uut For Calibration

    UUT Calibration Setup Connect the Calibration System to a UUT gtu087.jpg Figure 4-1. UUT Connection Procedure Prepare the UUT for Calibration Prepare the UUT for calibration (refer to the UUTs users documentation for instructions): 1. Read the UUT users documentation for information such as how to connect the inputs, turn-on inputs, configure the IP address and subnet-mask, assign or find the UUT ID number, and how to change the UUT configuration.
  • Page 76: Current Connections

    6135A/PMU Operators Manual Current Connections Connect the current leads from the Calibration System Input/Output Panel to the UUT as follows (refer to Figure 4-2). Note Most PMUs will have a visual current flow indicator (usually a “sense” or “phasing” dot) above the current input connectors. To prevent inaccurate UUT measurements, make sure to connect the red current wires to the current inputs with the current flow indicator.
  • Page 77: Voltage Connections

    UUT Calibration Setup Connect the Calibration System to a UUT Voltage Connections Connect the voltage leads from the Calibration System Input/Output Panel to the UUT as follows (refer to Figure 4-3): Note: The voltage leads supplied with the system provide 4 wire connection to the front panel of the 6135A should that be required for other applications.
  • Page 78: Uut Data Cable Connection

    6135A/PMU Operators Manual UUT Data Cable Connection Connect the network Ethernet cable from the Calibration System Input/Output Panel to the UUT as follows (refer to Figure 4-4): 1. On the rear panel of the Server PC, put one end of the Ethernet Cable into the left most Ethernet Port (see Figure 4-4).
  • Page 79: Communicate With The Uut

    3. Type in the Server IP address or Server PC name into the server address field. See Table 4-1 for more information on the Server Network Address selections. The default Server PC settings as supplied by Fluke are: IP address: 192.168.0.250 and PC name: PMUCal 4.
  • Page 80 “Modes of Operation” in Chapter 3. Manually type in the IP address of the Server PC or the Server PC name. The Server PC is supplied by Fluke with a fixed IP address of: 192.168.0.250 and a PC name of PMUCal. The entry in this field must reflect any changes to Server PC settings.
  • Page 81: Set Up A Uut Personality Profile

    UUT Calibration Setup Communicate with the UUT Set Up a UUT Personality Profile A UUT personality profile is a file that stores unique information about the UUT (refer to “UUT and Simulator Personality Profiles” in Chapter 3). The UUT personality profile contains information such as the UUT IP address, port number, ID, and the phasor order.
  • Page 82 Note The Fluke 6135A/PMU Calibration System can support Fluke 6105A/6106As with 50 A or 80 A options or up to three Fluke 52120A Transconductance Amplifiers per channel. The higher current outputs are taken directly from the units outputs. The Calibration Software...
  • Page 83 Settling Period Settling period in seconds before a measurement starts.  When a system contains a Fluke 6135A with either 50 A or 80 A options or has Fluke 52120A Transconductance Amplifiers connected, System Output a message appears in this area that informs the operator which ...
  • Page 84: Configure The Uut Phase Order

    6135A/PMU Operators Manual Note The System Software can support PMUs with inputs set to ratios other than 1:1. This allows PMUs which use input transducers, (transformers, current coils, and more), to be measured with their normal operating settings and without the need to modify these settings for testing. The UUT must be connected directly to the System without any input transducers during testing because they will affect the measurement.
  • Page 85 UUT Calibration Setup Communicate with the UUT Table 4-3. PMU Phase Order Screen gtu134.emf Item Name Description Calibrator Phase Calibration System phase names (Phase A, Phase B, Phase C, and the  Order List Positive Sequence inputs). Use to select which UUT Phase input corresponds with the Timing Unit System phases.
  • Page 86: Uut Connection Test

    6135A/PMU Operators Manual Table 4-3. PMU Phase Order Screen (cont.) Item Name Description Connect to PMU Connects to the PMU and returns the configuration. and Get PMU  Config Save Use to save the current UUT personality profile.  Load Use to select and load a saved UUT personality profile.
  • Page 87 UUT Calibration Setup UUT Connection Test 5. Plot voltage and current for Phases A, B , and C . Plot TVE, Phase Error , Magnitude Error, Frequency Error and ROCOF Error as shown on the Plot Configuration Dialog shown below. gtu150.png Figure 4-7.
  • Page 88 6135A/PMU Operators Manual In the Panel0 plot in Figure 4-8, Panel0 gives an indication of how the phase error should look. Note that the panels are self-scaling and the visual representation made be somewhat different. Check the vertical scale spans less than 0.5 degrees.
  • Page 89 UUT Calibration Setup UUT Connection Test In the Panel0 plot in Figure 4-9, the phase error for current phase B in Panel0 is 120 degrees the current phase C is -120 degrees. This indicates that these two phases are swapped. This could be a problem with the current wire connections but it could also be a mistake in the Phase order in the PMU Personality.
  • Page 90 6135A/PMU Operators Manual In Panel2 plot in Figure 4-10, Panel2 shows a Magnitude error of 70 volts on the phase b voltage channel. The TVE for phase B is 100%. These inaccuracies indicate that the voltage input is unconnected. gtu137.jpg Figure 4-10.
  • Page 91 UUT Calibration Setup UUT Connection Test In the Panel1 plot in Figure 4-11, Panel1 shows how the phase error should look. In the Panel1 plots in Figure 4-12, Panel1 shows approximately 100% TVE for all three current phases. The voltage phases are also 100% TVE. This could be an indication that the 6135A forced all the outputs off because of a problem with the voltage and current connections to the UUT.
  • Page 92 6135A/PMU Operators Manual gtu135.jpg Figure 4-12. Problem - All Outputs Forced Off 4-22...
  • Page 93: Automated Testing

    Chapter 5 Automated Testing Title Page About this Chapter ..................... 5-3 Automated Testing Overview ................5-3 Getting Started ....................5-3 Run an Automated Test ..................5-4 Incomplete Tests ....................5-6 Make a New Automated Test ................5-8...
  • Page 94 6135A/PMU Operators Manual...
  • Page 95: About This Chapter

    Automated Testing About this Chapter About this Chapter This Chapter supplies instructions on how to do an automated test on a UUT. All procedures in this chapter assume that a UUT is connected and configured. If the UUT is not connected and configured, refer to Chapter 4. For an alphabetical list of terms and abbreviations used in this chapter and manual, see “The Glossary”...
  • Page 96: Run An Automated Test

    6135A/PMU Operators Manual Run an Automated Test To run an automated test: 1. Use the instructions in Chapter 4 to prepare the UUT for test and calibration. 2. Click Test in the top Navigation Menu . 3. In General Interactive Testing , click Run Automated Test . A Windows Load dialog box opens (see Figure 5-1).
  • Page 97 Automated Testing Run an Automated Test gtu107.jpg Figure 5-2. Results File Selection 6. Monitor the test status screen as the automated test runs (see Figure 5-3). Note If “Skip Current Test” is clicked, the current test is stopped and the next test in the sequence is started.
  • Page 98: Incomplete Tests

    6135A/PMU Operators Manual Incomplete Tests After the automated test is complete, use the results log to complete incomplete tests. An incomplete test is defined as a test that has a status of “Cancelled”, “Skipped”, or “Error” on the log. To complete skipped automated tests: Note Skip to step 5 if you already have the applicable Results file loaded.
  • Page 99 Automated Testing Incomplete Tests 7. Click Run Incomplete Tests to run all of the incomplete tests. Notes After “Run all Incomplete Tests” is clicked, the PMUCal Software runs all the tests with an incomplete status of “Cancelled”, “Failed”, or “Error”. After the test is finished, the test statuses refresh.
  • Page 100: Make A New Automated Test

    6135A/PMU Operators Manual Make a New Automated Test The Calibration System comes with a full set of automated test files to fully calibrate a UUT to the limits in C37.118.1a - 2014. These files can be used as templates for special user defined test sequences.
  • Page 101 Automated Testing Make a New Automated Test Table 5-1. Setup Automated Test Window (cont.) Item Name Description Move Down Move the selected test down one position.  Delete Delete the selected test.  Clear List Clear the entire list.  Save Save the list.
  • Page 102 6135A/PMU Operators Manual 5-10...
  • Page 103: General Testing

    Chapter 6 General Testing Title Page About this Chapter ..................... 6-3 General Testing ....................6-3 Test Configuration Instructions ................. 6-6 General Test Parameters ................6-6 Test Type Parameter ..................6-7 Test Parameters ..................6-7 Steady State Test Parameters ..............6-8 Frequency Ramp Test Parameters .............
  • Page 104 6135A/PMU Operators Manual...
  • Page 105: About This Chapter

    General Testing About this Chapter About this Chapter This chapter supplies instructions on how to do an interactive test on a UUT. To do the procedures in this chapter, the UUT must be connected and configured. If the UUT is not connected and configured, refer to “Prepare the UUT for Calibration”...
  • Page 106 6135A/PMU Operators Manual Table 6-1. The Test Configuration Window gtu122.emf Item Name Description General Test Area that contains all the tests general test parameters (refer to “General Test  Parameters Area Parameters” in this chapter). Test Type Use to select the Test Type (refer to “Test Type Parameter” in this Chapter for ...
  • Page 107 General Testing General Testing Table 6-1. The Test Configuration Window (cont.) Item Name Description Simulate Use to run the selected test on the Simulated PMU.  Use to open a Save As dialog box that lets you save the current test. Note Save ...
  • Page 108: Test Configuration Instructions

    6135A/PMU Operators Manual Test Configuration Instructions Use this section to configure a custom interactive test. To configure a test, you must select a test type and set the general test parameters and also the UUT test parameters. General Test Parameters General test parameters specify the type of test and test timing parameters for each test file.
  • Page 109: Test Type Parameter

    General Testing Test Configuration Instructions Test Type Parameter The test type parameter is type of test that the test file is configured for. Select a Steady State, Frequency Ramp, Modulation, or Step Test (see “Standard Compliance Tests and Subtests” in Chapter 4). The test type parameter tells the PMUCal Software which test parameters in the test file to use when a test is run or simulated.
  • Page 110: Steady State Test Parameters

    6135A/PMU Operators Manual Steady State Test Parameters A Steady State test is where the magnitude, and frequency are constant for the duration of the test (see “Standard Compliance Tests and Test Files” in Chapter 3 for more information). See Table 6-3 for a detailed description of each parameter on the Steady State test window.
  • Page 111 General Testing Test Configuration Instructions Table 6-3. Steady State Test Configuration Window gtu125.emf Item Field Name Range Description Input Frequency (44.0 to 65.9) Hz Set the test frequency.  Set the phase offset at the test start. All (-180° to 180°) Phase Shift test signals are specified to be cosine ...
  • Page 112: Frequency Ramp Test Parameters

    6135A/PMU Operators Manual Frequency Ramp Test Parameters A Frequency Ramp test is a dynamic test that tests how the UUT measures changes in frequency (see “Standard Compliance Tests and Test Files” in Chapter 3 for more information). See Table 6-4 for a detailed description of each Frequency Ramp test parameter.
  • Page 113: Modulation Test Parameters

    General Testing Test Configuration Instructions Note The Start and End Frequencies are affected by Transition Time setting. Modulation Test Parameters A Modulation test is a dynamic test that tests UUT response to cosine modulated phase, amplitude, or both. (see “Standard Compliance Tests and Test Files” in Chapter 3 for more information).
  • Page 114 6135A/PMU Operators Manual Table 6-5. Modulation Test Configuration Window gtu127.emf Item Field Name Range Description Input Frequency (44.0 to 65.9) Hz Set the frequency for the test.  Set the phase offset at the start of the test. All test Phase Shift (-180°...
  • Page 115: Step Test Parameters

    General Testing Test Configuration Instructions Step Test Parameters A Step Test is a dynamic test that tests the UUTs response to steps in magnitude or phase. (see “Standard Compliance Tests and Test Files” in Chapter 3 for more information). See Table 6-6 for a detailed description of each step test Parameter.
  • Page 116 6135A/PMU Operators Manual Table 6-6. Step Test Configuration Window gtu128.emf Item Field Name Range Description Input Frequency (44.0 to 65.9) Hz Set the frequency for the test.  Set the phase offset at the start of the test. All test Phase Shift (-180°...
  • Page 117: Measurement Latency Test Parameters

    General Testing Run a Single Test Table 6-7. Latency Test Configuration Window gtu180.emf Item Field Name Description The Test Duration in seconds. The duration should Test Duration be long enough to collect a sufficient number of  measurements. Measurement Latency Test Parameters To configure a Measurement Latency Test (refer to Table 6-7).
  • Page 118: Run A Custom Interactive Test

    6135A/PMU Operators Manual Run a Custom Interactive Test 1. Click Test in the top Navigation Menu . 2. Click Customize and Run Single Test . A Windows Explorer dialog box opens. 3. Load a test file with similar test parameters (if desired): a.
  • Page 119: Interactive Graphs, Results, And Reports

    Chapter 7 Interactive Graphs, Results, and Reports Title Page Introduction ......................7-3 Interactive Graph ....................7-3 Graph Configuration Window ............... 7-4 Make an Interactive Graph ................7-4 Example Interactive Graph Configurations ........... 7-7 One Source and One Input Graph Configuration Example ....... 7-7 Two Sources and One Input Graph Configuration Example ....
  • Page 120 6135A/PMU Operators Manual...
  • Page 121: Introduction

    Interactive Graphs, Results, and Reports Introduction Introduction This chapter supplies the information necessary to make and read interactive graphs, test results, test reports, and certification reports. For an alphabetical list of terms and abbreviations used in this chapter and manual, see “The Glossary”...
  • Page 122: Graph Configuration Window

    6135A/PMU Operators Manual Graph Configuration Window To make and configure an interactive graph, use the graph configuration window (see Table 7-1). The graph configuration window automatically opens after the test data is collected and analyzed after a test is run.
  • Page 123 Interactive Graphs, Results, and Reports Interactive Graph Table 7-1. Interactive Graph Configuration Window gtu121.emf Item Name Description Use to select the source data to plot. The two selections are: • Reference (6135A Output) – Plot the test data from the 6135A System.
  • Page 124 6135A/PMU Operators Manual Table 7-1. Interactive Graph Configuration Window (cont.) Item Name Description Use to select the which input to plot. The two selections are: • Voltage – Plot the voltage test data from the source selected. • Current – Plot the current test data from the source selected.
  • Page 125: Example Interactive Graph Configurations

    Interactive Graphs, Results, and Reports Interactive Graph Table 7-1. Interactive Graph Configuration Window (cont.) Item Name Description Uses the selections and makes the interactive graph(s). This Plot Results  button remains inactive (greyed out) until sufficient selections are made to create at least one interactive plot. Save ...
  • Page 126: Two Source And Two Inputs Graph Configuration Example

    6135A/PMU Operators Manual Two Source and Two Inputs Graph Configuration Example Figure 7-5 is an example graph configuration with two sources and two inputs selected. Four windows opens when configured as shown. Source Actual Voltage Current Inputs to Plot gtu093.emf...
  • Page 127: Y-Axis Analysis Results Configuration Example

    Interactive Graphs, Results, and Reports Interactive Graph Y-Axis Analysis Results Configuration Example Figure 7-6 is an example graph configuration with all the analysis selections selected. Each selection will open a separate panel in the window. Y Axis: Analysis Results TVE (Percent) Magnitude Error Phase Error Frequency Error...
  • Page 128: Change The View Of An Interactive Graph

    6135A/PMU Operators Manual Change the View of an Interactive Graph Zoom-in and Zoom-out of a Plot To see the data closer, zoom-in and zoom-out of the plot. Zoom selection are: vertically, horizontally, and quadrant. To zoom-in and zoom-out (refer to Figure 7-7 and 7-8): 1.
  • Page 129: Undock And Dock A Panel

    Interactive Graphs, Results, and Reports Interactive Graph Click and drag to make a Shape of a Box a then Release gtu100.emf Figure 7-8. Example Quadrant Zoom Undock and Dock a Panel The panels in the interactive graph window can be undocked (removed from) and docked (put back into) to let you resize the panel to get a closer view of the data.
  • Page 130: Change The Displayed Data

    6135A/PMU Operators Manual Change the Displayed Data The data shown in the panel can be changed as necessary to see different data. To change the data shown in the panel (refer to Figure 7-10), left-click in on the combo box under Measurement , Domain , or Magnitude and select the new data to be shown.
  • Page 131: Measurement Latency Test Results

    Interactive Graphs, Results, and Reports Measurement Latency Test Results Measurement Latency Test Results The Measurement Latency Test presents the results as a table (see Figure 7-12). gtu181png Figure 7-12. Measurement Latency Test Results Table The Max Latency result is the measurement of interest in a Latency test. The other values, (Min, Mean and Standard Deviation), provide an indication of the spread of the values within the measurement epoch.
  • Page 132: Test Results

    6135A/PMU Operators Manual Test Results Test Results Viewer The test results viewer (Show Results button) is a spreadsheet view of the test results that shows maximum test results for each compliance test and subtest. The results viewer shows the test results in the active results file that is shown in the result file path in the top navigation menu.
  • Page 133: Log Of Tests

    Interactive Graphs, Results, and Reports Test Results Log of Tests The Show Log button on the results viewer screen shows a complete list of tests run and their completion status. From the Show Log screen (see Figure 7-14), tests can be directly modified then tested again.
  • Page 134: Test Reports

    6135A/PMU Operators Manual Test Reports From the results screen, a formal calibration report can be generated. To make a calibration report, click Results on the top navigation menu then select Create One Report or Create Multiple Reports. One report makes a single report with all test results.
  • Page 135: Create One Report

    Interactive Graphs, Results, and Reports Test Reports Create One Report To create a single report of all test results: 1. Click on Create One Report. A prompt will appear asking which report template to use. See Figure 7-15. 2. Select the report template required and click on OK. A report is now created using Microsoft Excel and on completion the location is reported in a dialog box.
  • Page 136 6135A/PMU Operators Manual created using Microsoft Excel and on completion, their location is reported in a dialog box. Individual reports can be viewed clicking on the View Existing Report. Note The time it takes to create a report is dependent on the complexity of the report template.
  • Page 137: View A Report

    Interactive Graphs, Results, and Reports Test Reports View a Report Individual reports can be viewed by clicking on View Existing Report. A screen will appear that will allow navigation to a report. See Figure 7-17. Select the required report and click on OK. Microsoft Excel will open displaying the report.
  • Page 138 6135A/PMU Operators Manual gtu120.emf Figure 7-18. Example Report 7-20...
  • Page 139: Troubleshooting, Maintenance, And Care

    Title Page About this Chapter ..................... 8-3 Troubleshooting ....................8-4 General ......................8-4 6135A/PMU System Timing Unit ..............8-4 Server PC ....................... 8-5 6135A Three Phase Power Standard ............. 8-5 GPS Receiver ....................8-6 Network ......................8-6 Calibration Software ..................8-6 UUT .......................
  • Page 140 6135A/PMU Operators Manual...
  • Page 141: About This Chapter

    Troubleshooting, Maintenance, and Care About this Chapter About this Chapter This chapter provides information to troubleshoot the Calibration System. For an alphabetical list of terms and abbreviations used in this chapter and manual, see “The Glossary” in Appendix A of this manual. For system connection information, see Figure 8-1.
  • Page 142: Troubleshooting

    4. Power module or power circuit 2. Unplug the unit and check the board is bad. fuses. 6135A/PMU System Timing Unit For information on the LEDs on the front panel, see “6135A/PMU System Timing Unit” in Chapter 1. Problem Probable Cause Action...
  • Page 143: Server Pc

    AUX Indicator LED flashes. Lost connection to Server PC. Restart the Server PC. Server PC For information on the LEDs on the front panel, see “6135A/PMU System Timing Unit” in Chapter 1. Problem Probable Cause Action POWER LED is off when power is When the LED is off, the PC is not See the Server PC Users Manual.
  • Page 144: Gps Receiver

    Complete these steps in order: four satellites. 1. Make sure the 6135A/PMU System Timing Unit is on. 2. Antenna could be 2. Check the antenna position. See the Microsemi GPS damaged or dirty. Receiver and Antenna Installation Manual for troubleshooting procedures.
  • Page 145: Uut

    Troubleshooting, Maintenance, and Care Troubleshooting Problem Probable Cause Action This error message appears: 1. The Server PC is not Complete these steps in order: connected to the UUT. 1. Make sure the UUT is on. 2. Incorrect UUT IP address or 2.
  • Page 146 GPS is locked to a the GPS system. 3. On the rear panel of the 6135A/PMU System Timing Unit, disconnect the “GPS Antenna In” and connect it directly to the “GPS In” on the UUT.
  • Page 147: Test Reports

    Lo of the failing Current Phase. Test failures on a Current Phase cannot be caused by open circuit current connections as the 6135A/PMU Calibrator would report voltage over compliance errors and the system would stop. Failures on one Current...
  • Page 148: Cleaning

    Performance Verification and Repair Information Annual performance verification and adjustment is recommended to verify performance to the Calibration System specifications. Fluke Calibration is the only Service Center that can adjust and repair a Calibration System. To schedule and send the Calibration System to Fluke for calibration or repair: 1.
  • Page 149: Performance Verification

    Chapter 9 Performance Verification Title Page About this Chapter ..................... 9-3 Prerequisites ....................... 9-3 Equipment Required ..................9-3 Procedures ......................9-3 Verify Signal Alignment with the UTC ............9-3 Verify the Accuracy of the Timing Inputs ............. 9-4 Verify Step Timing Accuracy................ 9-5 L1V Test ....................
  • Page 150 6135A/PMU Operators Manual...
  • Page 151: About This Chapter

    Performance Verification About this Chapter About this Chapter This chapter provides information on how to verify the performance of the Calibration System. Prerequisites 1. A recently calibrated 6105A Master Unit, and 6106A Auxiliary Units 1 and 2 installed in the Calibration System. 2.
  • Page 152: Verify The Accuracy Of The Timing Inputs

    This test measures the accuracy of the time inputs STU_T0 and STU_TRefp data input to the mathematical model. 1. Open the file: <userprofile>\Documents\Fluke PMUCal\System\Logs\T0 time stamp.txt. 2. The value “Diff” (3 parameter) is in seconds. Record Diff in the “Model time difference”...
  • Page 153: Verify Step Timing Accuracy

    Performance Verification Procedures Verify Step Timing Accuracy This test checks the latency of the step change in the 6135A analogue output with respect to the 1 PPS. The objective is to compare the time of the step with respect to the 1 PPS with the time stamp STU_T0.
  • Page 154: L1V Test

    6135A/PMU Operators Manual L1V Test 1. Connect channel 1 of the oscilloscope input to the L1V connection on the Calibration System Input/Output panel (right side of the system). 2. In the Test Parameters menu set: • Input Frequency: to nominal frequency (50 Hz or 60 Hz) •...
  • Page 155: L2V Test

    4. When the oscilloscope has triggered, use the vertical cursors to measure and record in Table 9-4 the values of Ta, Tb, and Tc for L2V as shown as shown in Figure 9-1. 5. Open the file: 6. <userprofile>\Documents\Fluke PMUCal\System\Logs\T0 time stamp.txt The value “Diff” (3...
  • Page 156: L3V Test

    4. When the oscilloscope has triggered, use the vertical cursors to measure and record in Table 9-4 the values of Ta, Tb, and Tc for L3V as shown as shown in Figure 9-1. 5. Open the file: <userprofile>\Documents\Fluke PMUCal\System\Logs\T0 time stamp.txt. 6. The value “Diff” (3 parameter) is in seconds.
  • Page 157: A Glossary

    Comma-separated values file. Each line is one entry or record and the fields in a record are separated by commas. Calibration Software Fluke Calibration Software used to Calibrate a UUT. Default Test Files Test files that are made with by the Calibration Software when the software is configured for use.
  • Page 158 6135A/PMU Operators Manual Harmonics A waveform that is an integral multiple of the fundamental frequency. For example, a waveform that is twice the frequency of a fundamental is called the second harmonic. Interactive Test Testing method that requires the user to configure, run, and evaluate the test data.
  • Page 159 Glossary Introduction ROCOF Rate of Change of Frequency or “ROCOF” is the rate that the frequency changes. Step Tests See C37.118.1 for definitions of Response Time, Delay Time and Overshoot. Synchrophasor A Phasor measurement that has a time stamp is referred to as a “Synchrophasor”. Testing Methods Type of test to be performed on the UUT.
  • Page 160 6135A/PMU Operators Manual...

This manual is also suitable for:

6135a/pmucal

Table of Contents