Basic Operation

Select a Function

After the power is switched on and the initial checks are complete, the Product
resets to its reference state:
Function:
Voltage:
Current:
Phase:
Frequency:
3-10
Table 3-6. Rear Panel (cont.)
Power line voltage selector (115 V or 230 V settings)
Fuse for internal multimeter (T100 mA / 250 V)
Central ground terminal
GPIB connector
Phase reference output: Internal reference signal for 0 ° phase. All output
channels are synchronized with this signal.
Sample reference output: Used to trigger an external voltmeter to synchronize
sampling measurements with the Instrument outputs, useful to calibrate the
Product. Enabled only by remote control.
Energy pulse output (active with 6003A/E Energy Option): Energy pulse output
from the Product. Frequency of pulses is proportional to the power generated by
the Product and is defined by the energy meter constant. See Calibrator Setup
Menu and Generating Electric Energy in Chapter 4.
USB connector
Pac Basic (AC Power, Basic mode)
10 V (Channel 1 active)
1 A (Channel 1 active)
0 °
50 Hz
Explanation