Set time interval for auto inversion mode.
Setting range: 5-999 sec
Default setting is "5"
1. Enter user setting
Fwd
Rev
Press 2 sec
2. Change auto inversion time
setting
Fwd
Rev
Fwd
Rev
3. Exit user setting
Fwd
Rev
① Press the Power key.
AutoR
Sync
① Turn the Encoder dial and select "SEC".
② Press the Encoder dial.
③ Turn the Encoder dial to set desired time
AutoR
Sync
④ Press the Encoder dial to finalize.
AutoR
Sync
After completing the setting, press the Encoder dial for
two seconds. Display reverts to previous screen and
shows current rpm.
AutoR
Sync
Press 2 sec
4. OPERATION PROCEDURES
Press the Encoder dial for two seconds while current
rpm is on the screen.
Unit enters user setting.
RPM display: "SEC" flashes
RPM display: Current setting flashes
Setting range: 5-999 sec