Make connection setting with vacuum controller VR102 (sold separately).
OFF: Not linked (does not carry out communication)
ON: Linked (carries out communication)
Default setting is "OFF"
While connection is "ON", RE unit is interlocked with vacuum controller in starting or stopping
operation.
1. Enter user setting
Fwd
Rev
Press 2 sec
2. Change connection setting
Fwd
Rev
Fwd
Rev
3. Exit user setting
Fwd
Rev
AutoR
Sync
AutoR
Sync
AutoR
Sync
AutoR
Sync
Press 2 sec
4. OPERATION PROCEDURES

Vacuum Controller Connection

① Turn power ON (|).
② Press the Encoder dial for two seconds while current
rpm is on the screen.
Unit enters user setting.
③ Turn the Encoder dial and select "di".
RPM display: "di" flashes
④ Press the Encoder dial.
RPM display: Current setting flashes
⑤ Turn the Encoder dial to set select ON/OFF.
⑥ Press the Encoder dial to finalize.
After completing the setting, press the Encoder dial for
two seconds. Display reverts to previous screen and
shows current rpm.