Page 1
DP43C DP63C - First Edition - Thank you for purchasing " Vaccum Drying Oven, DPC Series" of Yamato Scientific Co., Ltd. To use this unit properly, read this "Instruction Manual" thoroughly before using this unit. Keep this instruction manual around this unit for referring at anytime.
Contents Cautions in Using with Safety..............1 • Explanation........................1 • Table of Illustrated Symbols ..................2 • Fundamental Matters of "WARNING!" and "CAUTION!" ..........3 Before Using this unit ................4 • Requirements for Installation..................4 Description and Function of Each Part ..........9 •...
Cautions in Using with Safety Explanation MEANING OF ILLUSTRATED SYMBOLS Illustrated Symbols Various symbols are used in this safety manual in order to use the unit without danger of injury and damage of the unit. A list of problems caused by ignoring the warnings and improper handling is divided as shown below.Be sure that you understand the warnings and cautions in this manual before operating the unit.
Cautions in Using with Safety Table of Illustrated Symbols Warning Warning, Warning, Warning, Warning, Warning, generally high voltage high temperature drive train explosive Caution Caution, Caution, Caution, Caution, Caution, generally electrical shock scald no road heating not to drench Caution, Caution, water only deadly poison...
Cautions in Using with Safety Fundamental Matters of "WARNING!" and "CAUTION!" WARNING! Do not use this unit in an area where there is flammable or explosive gas Never use this unit in an area where there is flammable or explosive gas. This unit is not explosion-proof.
WARNING! 1. Always ground this unit • The DP23C/33C type use a single-phase 115V, DP43C/63C types use a single-phase 220V power source. Please consult your local electrical contractor for power connecting work. • Be sure to connect the earth wire (the green cable of power cord) to the grounding conductor or ground terminal to prevent accidents caused by electric leakage.
Page 7
Before Using this unit Requirements for Installation 3. Do not use this unit in an area where there is flammable or explosive gas (Refer to page 44 "List of Dangerous Substances".) • Never use this unit in an area where DP43C Vacuum Drying Oven there is flammable or explosive gas.
Page 8
Before Using this unit Requirements for Installation 4. Do not modify 5. Installation on horizontal surface • Modification of this unit is strictly • Set this unit to the flattest place. Setting prohibited. This could cause a failure. this unit on rough or slope place could cause the vibration or noise, or cause the unexpectible trouble or malfunction.
Page 9
• Choose a correct power distribution board or receptacle that meets the unit’s rated electric capacity. Electric capacity: DP33C: 220V AC, 4.8A DP23C: 220V AC, 3.1A DP63C: 220V AC, 14.3A DP43C: 220V AC, 10.3A NOTE) There could be the case that the unit does not run even after turning ON the power. Inspect whether the voltage of the main power is lowered than the specified value, or whether other device(s) uses the same power line of this unit.
Page 10
Modle of suction Maker Suitable Necessary plumbing part Modle of port of vacuum pump vaccum oven vacuum pump DP23C/33C PD(PX)52 Dia.18mm Vacuum hose:I.D.15mm 1pc (Nipple of vacuum oven Vacuum hose: I.D.15mm 1pc Yamato for connecting PD(PX)102 Vacuum hose: I.D.25mm 1pc Dia.30mm...
Description and Function of Each Part Main Unit DP43C/63C Door Observation window Caution plate Door handle Cable port Intake speed regulating valve Exhaust speed (Main c ircuit breaker ) regulating valve Caster Operation panel Vacuum Guage...
Description and Function of Each Part Control Panel ⑧ ⑭ ⑨ ⑮ ⑩ ⑯ ⑪ ⑫ ① ⑬ ③ ② ⑦ ④ ⑥ ⑤ ① START/STOP Key: Starts/stops the operation. ② ▲▼ Key: Uses for rising UP/lowering DOWN the setting value. ③...
Description and Function of Each Part Characters of the Controller The characters VS4 controller shows are as follows: Character Identifier Name Purpose Fixed Temperature Used for starting the fixed temperature Setting Mode operation. Temperature Setting Used for setting the temperature. Timer Setting Mode Represents the setting of quick auto stop or AStP...
Operation Method Operation Mode and Function List All the operation mode of this unit is as follows; Name Description Page Pressing the FIXED TEMP key enters into the fixed temperature operation setting mode. Fixed Temperature Pressing it again enters into the temperature setting mode. Operation The "▲▼"...
Page 15
Operation Method Operation Mode and Function List The operation function of this unit is as follows; Name Description Page This function is set to be automatically activated Auto overheating (auto reset) when the temperature exceeds the prevention function setting temperature by 12℃. Though the device shares power source, display, Overheating and key input with the controller, it has independent...
Operation Method Operation Mode, Function Setting Key, and Characters The operation mode setting and function setting use the key operation and characters show in the following figure.
To prevent such matters, set a proper value. The DP23C/33C temperature is set to 260℃ at factory shipment (The DP43C/63C temperature is220℃). Do not set the value larger than it.
Operation Method Fixed Temperature Operation 1. Turn on the power (turn on the breaker in front) Fixed temperature operation procedure The default value is displayed for about four seconds after turning on the power. The screen then displays the initial setting. The current temperature in furnace, operation mode character and setting temperature of overheating prevention device are displayed on respective screens.
Operation Method Quick Auto Stop Operation This operation is used to specify the period up to automatic stop, i.e., Quick auto stop sets the auto stop timer during operation. operation procedure 1. Set the time up to stop during fixed temperature operation •...
Operation Method Auto Stop Operation This operation is used to specify the automatic stop time in the fixed Auto stop operation temperature operation. procedure 1. Set stop time ① Press the TIMER key on the initial screen. Press the TIMER key again. The setting temperature display screen displays the character "AstP", which indicates the auto stop operation, with blinking.
Page 21
Operation Method Auto Stop Operation 3. Stop/terminate timer operation • The operation stops automatically at setting time. • Buzzer continues to sound for about five minutes at operation stop. • The setting temperature screen displays the character "End", which indicates termination of operation, with the FIXED TEMP and AUTO STOP lamps lighting on.
Operation Method Auto Start Operation This operation is used to specify the period up to automatic start after Auto start operation power on. procedure 1. Set start time ① Press the TIMER key on the initial screen. Press the TIMER key again. The setting temperature display screen displays the character "Astr", which indicates the auto start operation, with blinking.
Page 23
Operation Method Auto Start Operation 3. Stop/terminate timer operation • The operation starts automatically at setting time. • Press the START/STOP key for one second to stop or terminate operation. The screen returns to the initial setting screen. To correct or check setting… Changing the setting temperature or time during operation is possible by pressing the TIMER key.
100℃ to 150℃ for DP33C). Temperature stability time after reaching to the setting temperature is necessary to be added. Make sure to conduct a test run before setting the optimum time. Condition: room temperature 23℃, no load, exhaust opening fully opened (unit: minute) DP23C DP33C DP43C DP63C...
Page 25
Operation Method Program Operation Program creation The program pattern below is explained as an example. 1. Program pattern example Temp. 200℃ 100℃ 0℃ Step ← → Repeat function Temp.(℃) The number of steps is not counted. Time(min.) 1. Turn on the power •...
Page 26
For example, about 60 minutes is needed to increase the temperature from room temperature to 240 ℃ for DP23C type (90minutes for DP33C type). When the current temperature is supposed to be 25℃, it takes about 3 minutes to increase the temperature by 1 ℃ , accordingly takes about 60 minute until it reaches to 240℃.
Page 27
Operation Method Program Operation 5. Start program operation • Press the [START/STOP] key for about one second. program operation previously set starts. • The PROGRAM lamp lights on and the setting temperature screen displays the step currently under operation. Press the "▼" to check the setting temperature and residual time of step currently under operation on the setting temperature screen.
Page 28
Operation Method Program Operation Use program repeat function This section explains how to register the program repeat (repeating a program pattern) in program operation. This section explains the registration procedure of program using repeat function in "4. Register program" above. The procedure sets the step number to be repeated "PS-n"...
Page 29
Operation Method Program Operation Programming Preparation Form 1 (Please use this form by making copies) Register with: PrG1 PrG2 PrG3 PAt1 PAt2 PAt3 Date Project Name Programmer Program Pattern 250℃ 200℃ 150℃ 100℃ 50℃ STEP...
Page 30
Operation Method Program Operation Programming Preparation Form 2 (Please use this form by making copies) Register with: PrG1 PrG2 PrG3 PAt1 PAt2 PAt3 Date Project Name Programmer Input Value Temperature (℃) Time (min.) Repeat Function Step 1 To/Times Step 2 Step 3 Step 4 Step 5...
Operation Method Other Functions Use calibration offset function Calibration offset is a function which corrects the difference between the temperature in furnace and that of controller (sensor temperature) if arises. The function parallel corrects the difference either to the plus or minus side within the whole temperature range of unit. The function can be set or cancelled by the SUBMENU key.
Operation Method Vacuumizing and air intake Vacuumizing: exhaust Speed 1) Turn the intake speed regulating valve CW to its end (close); 2) Turn the exhaust speed regulating valve CCW to its end, the exhaust speed is max. here; 3) Turn on vacuum pump for vacuumizing. Air intake: Turn the exhaust speed regulating valve CW to its end (close);...
Do not use the sample includes those. Use under proper temperature range Operational temperature range of DP23C/33C/is room temperature +5 to 240℃. Operational temperature range of DP43C/63C is room temperature +5 to 200℃. Never set the temperature out of that.
Page 34
Handling Precautions Setting of sample Since the withstand load of the attached shelf plate is about 15kg per one plate, do not set heavier sample than 15kg. When setting several sample, set them as dispersed as possible. Too much sample setting could cause the improper control of the temperature. For keeping the proper temperature, keep more than 30% space against whole size of the shelf plate, and set the sample.
Maintenance Method Daily Inspection and Maintenance For the safety use of this unit, please perform the daily inspection and maintenance without fail. Using the city water to this unit might attach dirt. Do inspect and maintain this point while performing daily inspection and maintenance.
Long storage and disposal When not using this unit for long term / When disposing CAUTION! When not using this unit for long term… • Turn off the power and disconnect the power cord. WARNING! When disposing… • Keep out of reach of children. •...
In the Event of Failure… Safety Device and Error Code This unit has an automatic diagnosis function built in the controller and safety devices independent of the controller. The table below shows the cause and the solution method when the safety device operates. Error Code: When an abnormal condition occurs, an error code appears and the alarm lamp lights in the controller, the buzzer sounds simultaneously.
The power supply voltage is lower than the proper value. In the case if the error other than listed above occurred, turn off the power switch and primary power source immediately. Contact the shop of your purchase or nearest Yamato Scientific Service Office.
If the failure occurs, stop the operation, turn OFF the power switch, and unplug the power plug. Please contact the sales agency that this unit was purchased, or the Yamato Scientific's sales office. < Check following items before contact >...
Specification DP23C DP33C DP43C DP63C Method Pressure relief and wall heating Temperature control Room temp +5℃. to 240℃ Room temp +5℃. to 200℃ range Temperature ±1.5℃ (Middle of furnace, Set temp.: ±1℃ (Middle of furnace, Set temp.: adjustment 240℃,) 200℃,) accuracy Room temp.
Wiring Diagram DP23C/33C/43C/63C TM 1 AC220V PI O CO NT Symbol Part name Symbol Part name Main circuit breaker CONT Control board Terminal block Display circuit board Heater Sensor for control Main relay Sensor for overheating prevention Solid state relay...
Reference Capability Guidline DPC temperature rising curve Furnace center Furnace center Temp senso Temp senso time(min) DP23C temperature rising characteristics DP33C temperature rising characteristics Furnace center Furnace center Temp senso Temp senso 120 180 240 300 time(min) time(min) DP23C temperature rising characteristics...
Reference List of Dangerous Substances Never use explosive substances, flammable substances and substances that include explosive or flammable ingredients in this unit. EXPLOSIVE Ethylene glycol dinitrate (nitro glycol), Glycerin trinitrate (nitroglycerine), Cellulose nitrate (nitrocellulose), and other explosive nitrate esters Trinitrobenzene, Trinitrotoluene, Trinitrophenol (picric acid), and other explosive EXPLOSIVE: nitro compounds Acetyl hidroperoxide (peracetic acid), Methyl ethyl ketone peroxide, Benzyl...
Page 47
Responsibility Please follow the instructions in this document when using this unit. Yamato Scientific has no responsibility for the accidents or breakdown of device if it is used with a failure to comply. Never conduct what this document forbids. Unexpected accidents or breakdown may result in.