- First Edition - Thank you for purchasing "Forced Convection Constant Temperature Oven, DNE Series" of Yamato Scientific Co., Ltd. To use this unit properly, read this "Instruction Manual" thoroughly before using this unit. Keep this instruction manual around this unit for referring at anytime.
Contents Cautions in Using with Safety..............1 Explanation........................1 Table of Illustrated Symbols ..................2 Fundamental Matters of "WARNING!" and "CAUTION!" ..........3 Before Using this unit ................4 Requirements for Installation ..................4 Description and Function of Each Part..........9 Main Unit ........................9 Control Panel......................12 Operation Method...................
Page 3
Handling Precautions ................55 Maintenance Method................56 Daily Inspection and Maintenance ................56 Long storage and disposal ..............57 When not using this unit for long term / When disposing..........57 In the Event of Failure… ................ 58 Safety Device and Error Code ...................58 Trouble Shooting .......................60 After Service and Warranty ..............
Cautions in Using with Safety Explanation MEANING OF ILLUSTRATED SYMBOLS Illustrated Symbols Various symbols are used in this safety manual in order to use the unit without danger of injury and damage of the unit. A list of problems caused by ignoring the warnings and improper handling is divided as shown below.Be sure that you understand the warnings and cautions in this manual before operating the unit.
Cautions in Using with Safety Table of Illustrated Symbols Warning Warning, Warning, Warning, Warning, Warning, generally high voltage high temperature drive train explosive Caution Caution, Caution, Caution, Caution, Caution, generally electrical shock scald no road heating not to drench Caution, Caution, water only deadly poison...
Cautions in Using with Safety Fundamental Matters of "WARNING!" and "CAUTION!" WARNING! Do not use this unit in an area where there is flammable or explosive gas Never use this unit in an area where there is flammable or explosive gas. This unit is not explosion-proof.
Before Using this unit Requirements for Installation WARNING! 1. Always ground this unit The DNE400 and DNE600 types use a 115V power source. The DNE410, DNE610, DNE810 and DNE910 types use a single-phase 220V power source. Please consult your local electrical contractor for power connecting work. Be sure to connect the earth wire (the green cable of power cord) to the grounding conductor or ground terminal to prevent accidents caused by electric leakage.
Page 8
Before Using this unit Requirements for Installation 3. Do not use this unit in an area where there is flammable or explosive gas (Refer to page 69 "List of Dangerous Substances".) Never use this unit in an area where there is flammable or explosive gas. This unit is not explosion-proof.
Page 9
Before Using this unit Requirements for Installation 4. Do not modify 5. Installation on horizontal surface Modification of this unit is strictly Set this unit to the flattest place. Setting prohibited. This could cause a failure. this unit on rough or slope place could cause the vibration or noise, or cause the unexpectible trouble or malfunction.
Page 10
Before Using this unit Requirements for Installation 8. Do not use corrosive sample Stainless steel SUS304 is used for the main hot-air path; however, it may be corroded by strong acid etc. And the door packing made of silicon rubber may be corroded by some kind of solvent, e.g.
Page 11
Before Using this unit Requirements for Installation 12. Setting of the shelf and sample The number of shelves attached varies depending on the type of product (2 to 8).One of them (two for DNE910) is previously fixed on the lowermost stand of bracket supporter with screws at factory shipment.
Description and Function of Each Part Main Unit DNE400/410/600/610 Front view Exhaust openings Production plate Cable port Door Shelf Door handle Intake opening Specification plate Overheating prevention device Earth leakage breaker Control panel Rear view Optional accessory attaching port Power code (DNE410/610) Power code (DNE400/600)
Page 13
Description and Function of Each Part Main Unit DNE810 Front view Control panel Overheating prevention device Earth leakage breaker Production plate Cable port Door Intake opening Door handle Specification plate Rear view Optional accessory attaching port Exhaust openings Power code...
Page 14
Description and Function of Each Part Main Unit DNE910 Front view Control panel Overheating prevention device Earth leakage breaker Production plate Cable port Door handle Intake opening Door Specification plate Rear view Optional accessory attaching port Exhaust openings Power code...
Description and Function of Each Part Control Panel FIXED TEMP DOOR AUTO STOP ℃ SUBMENU AUTO START TROUBLE PROGRAM STANDBY POWER HEATER FUNCTION PROGRAM MENU ENTER CANCEL Displays the measured temperature and error code. Main Display: Displays the operation and setting information. Sub Display:...
Operation Method Operation Mode and Function List The operation mode consists of the following four modes. Name Description Page Controls temperature with fixed temperature. Fixed Temperature Operation Stops operation at specified time. Auto Stop Operation Starts operation at specified time. Auto Start Operation Starts program operation at specified time.
Operation Method Fixed Temperature Operation Start the operation from turning on the power shown in the figure, and continue the operation under the setting temperature unless turning off the power. SV: Setting temperature t: Time ▲ ▲ STOP START Setting of the Fixed Temperature Operation Turn on the power Turn on the power switch of the unit (earth leakage breaker).
Page 19
Operation Method Fixed Temperature Operation Set temperature ① The setting temperature input screen displayed. The Sub Display indicates "Set TEMP" and the numeric character that indicates FIXED TEMP DOOR Set TEMP 0℃ AUTO STOP SUBMENU ℃ AUTO START Fixed TEMP OPR temperature blinks.
Operation Method Auto Stop Operation As shown in the following figure, the device stops operating automatically by setting the timer. Set timer ▼ SV: Setting temperature t: Time Waiting Operation period Activate timer Start operation (manual) Stop operation (auto) Setting of the Auto Stop Operation Turn on the power Turn on the power switch of the unit (earth leakage breaker).
Page 21
Operation Method Auto Stop Operation Set temperature, operation period/stop ① The setting temperature input screen time, and fan displayed. The Sub Display indicates "Set TEMP" and the numeric character that indicates temperature blinks. FIXED TEMP DOOR Set TEMP 0℃ AUTO STOP SUBMENU ℃...
Page 22
Operation Method Auto Stop Operation Start operation ① Press the ENTER key to decide the setting and the auto stop operation starts. The blinking AUTO STOP lamp lights on and the Sub Display FIXED TEMP DOOR Set TEMP 100℃↑ AUTO STOP SUBMENU ℃...
Operation Method Auto Start Operation As shown in the following figure, this mode is applied to the device for starting the operation after the specified time (hours) automatically. Note that the device does not stop the operation automatically. Stop the operation by manual without fail. SV: Setting temperature Set timer t: Time...
Page 24
Operation Method Auto Start Operation Set temperature and start wait period/time ① The setting temperature input screen displayed. The Sub Display indicates "Set TEMP" and the numeric character that indicates FIXED TEMP DOOR Set TEMP 0℃ AUTO STOP SUBMENU ℃ ...
Operation Method Program Operation As shown in the figure, this mode is used for running the device under the setting program. Start pv (Temperature) Setting temperature Step1 Step2 Step3 Step4 Step5 Step6 Step7 Step8 (Time) Start program End program Setting of the Program Operation Turn on the power Turn on the power switch of the unit (earth leakage breaker).
Page 26
Operation Method Program Operation Set step number and start wait period/ ① The initiating step input screen is displayed. time The Sub Display displays "First step" and the step number blinks. FIXED TEMP DOOR ② Select the step number using the"▼▲" and then First step ...
Page 27
Operation Method Program Operation Start operation ① Press the ENTER key to decide the operation start wait period/time. The unit enters to program operation wait state. The blinking FIXED TEMP DOOR Step S01 AUTO STOP SUBMENU ℃ AUTO START Start in 30min PROGRAM lamp lights on and the STANDBY...
Operation Method Input Program Select program function ① Press the PROGRAM key. The program menu Check that the power in turned on. starts and step number selection screen is displayed. Press the "▼▲". The registered step numbers PROGRAM and the smallest number of un-used step are displayed in sequence.
Page 29
Operation Method Input Program Edit step Item Sub Display Notes Delete the step currently referred to. Delete step Step deletion sequence number of each step hereafter decreases by one. Program End Program end Complete program registration/edition. The step number currently edited is displayed on the lower column of Sub Display. The details for all registered steps in use are displayed followed by all un-used steps.
Operation Method Input Program; Set the Temperature Temperature setting procedure ① Select the "Set TEMP" and press the ENTER key. The temperature setting edition screen is displayed. The blinking " " (cursor) goes out and the setting value blinks instead. ►...
Operation Method Input Program; Set the Time Time setting procedure ① Select the "Set time" and press the ENTER key. The time setting edition screen is displayed. The blinking " " (cursor) goes out and the setting value blinks instead. ►...
Operation Method Input Program; Set the Fan Rotation Speed Fan rotation speed setting procedure ① Select the "Fan" and press the ENTER key. The setting edition screen is displayed. blinking " " (cursor) goes out and the setting value blinks instead. When "End"...
Operation Method Input Program; Set the Repeat Initiating Step Repeat initiating step setting procedure ① Select the "Rep. start" and press the ENTER key. The setting edition screen is displayed. The blinking " " (cursor) goes out and the setting value blinks instead. ►...
Operation Method Input Program; Set the Repeat Count Repeat count setting procedure ① Select the "Rep. count" and press the ENTER key. The setting edition screen is displayed. The blinking " " (cursor) goes out and the setting value blinks instead. ►...
Operation Method Input Program; Insert Step Step insertion procedure ① Select the "Insert step" and then press the ENTER key. A step is added there. ►0℃ Set TEMP Insert step ENTER → (Selection screen) (Selection screen for step inserted) The newly added step is inserted into the step number which is displayed on the selection screen. The sequence number of each registered step hereafter increases according to the number of steps added.
Operation Method Input Program; Add Step Step addition procedure ① Display the setting of un-used step on the setting item selection screen and then press the ENTER key. DOOR Set TEMP 0℃ S23(un-used) TROUBLE ENTER CANCEL The "SO1" (the first step) indicates "un-used" at the first registration of program. The items other than "Insert step", "Delete step"...
Operation Method Input Program; Delete Step Step deletion procedure ① Press the program key. The program menu starts and step number selection screen is displayed. Press the "▼▲". The registered step numbers and the smallest number of un-used step are displayed in sequence.
Operation Method Input Program; End Program End program procedure ① Select the "Program End" and press the ENTER key. The program edition is completed. DOOR Program End Standby TROUBLE → 2002/01/01 12:00 ENTER CANCEL ② The program edition has not done if the POWER or CANCEL key is pressed before pressing the "Program End"...
Operation Method Program Creation Method The program pattern below is explained as an example. - Program pattern example - Temp 200℃ 100℃ 0℃ Step Temp.(℃) Time(min.) 2:30 ← → Repeat function Repeat the steps S01 to S07 once Fan speed Rep.
Page 40
Operation Method Program Creation Method Edit step The unit operates with fixed rate inclination if the "START TEMP" and the "Set Temp" are different. It enters into "Wait" if the temperature does not reach to the setting temperature within a setting time.
Page 41
Operation Method Program Creation Method Edit step Refer to the page 54 for the temperature rise/fall time. They vary depending on the sample used or condition of exhaust opening (open or close). Check them by conducting operation. The conception of repeat is shown in the figure below. The first operation of repeat interval is not counted as a repeat count.
Page 42
Operation Method Program Creation Method Program creation; Step 1 (S01) ⑥ Press the PROGRAM key. The fan rotation speed ► 10 selection screen is displayed. The screen is not displayed when the wind speed adjustable device is not equipped. ⑦ Press the ENTER key. The fan rotation speed edition screen is displayed.
Page 43
Operation Method Program Creation Method Program creation; Step 3 (S03) to Step 7 (S07) ① Keep pressing the PROGRAM key until "SteP 3" (S03) is ► 0℃ Set TEMP displayed. S03(un-used) ② Edit in the same way to Step 6 (S06) and then keep DOOR pressing the PROGRAM key until the Step 7 (S07) is Set TEMP...
Page 44
Operation Method Program Creation Method Program creation; Step 8 (S08) to Step 9 (S09) ① Keep pressing the PROGRAM key until "SteP 8" (S08) is displayed. DOOR Set TEMP 0℃ ② Edit in the same way for Step 8 (S08). S09(un-used) TROUBLE ③...
Operation Method Set the Timer Mode Changes in timer mode are not reflected in the operation currently carried out in the following state of unit; operation start waiting state in the auto start mode, during auto stop mode operation, and operation start waiting state in the program operation mode.
Operation Method Set the Key Lock Mode Select the item in function menu Check that the power in turned on. ① Press the FUNCTION key. The function menu FUNCTION starts and the Sub Display displays the items. Select the item using the FUNCTION key. ②...
Operation Method Set the Buzzer Mode Select the item in function menu Check that the power in turned on. ① Press the FUNCTION key. The function menu FUNCTION starts and the Sub Display displays the items. Select the item using the FUNCTION key. ②...
Operation Method Calibration Offset Function Calibration offset is a function which corrects the difference between the temperature in furnace and that of controller (sensor temperature) if arises. The function parallel corrects the difference either to the plus or minus side within the whole temperature range of unit. 110℃...
Operation Method Integrating Operation Time The integrating operation time is a function to check the operation hours of unit. Its content can not be changed. Select the item in function menu Check that the power in turned on. ① Press the FUNCTION key. The function menu FUNCTION starts and the Sub Display displays the items.
Operation Method Set Clock The clock is not set at factory shipment. Set it with your watch or time tone before using the unit. The setting for clock can not be changed at the operation start waiting state in the auto start mode/program mode, or during operation.
Page 52
Operation Method Set Clock Input value ① The " " (cursor) goes out and the current value Date 2001/09/22 blinks instead. Display the desired value using Time 17:30 the "▼▲". ↓ Date 2002/09/22 Time 17:30 Set clock ① Press the ENTER key. The setting is decided and the function menu selection screen is DOOR Date...
Operation Method Set Fan Operation at Wait State Select the item in function menu Check that the power in turned on. ① Press the FUNCTION key. The function menu FUNCTION starts and the Sub Display displays the items. Select the item using the FUNCTION key. ②...
Operation Method Set the Fan Rotation Speed (Optional accessory) The setting value of fan rotation in each operation mode is displayed in the following three conditions. They, however, can not be changed. ・Operation completed state in auto stop mode, and the fan function parameter is set to "OFF". ・During operation or operation completed state in program operation mode.
Operation Method Set the Communication Lockout Mode (Optional accessory) Select the item in function menu Check that the power in turned on. ① Press the FUNCTION key. The function menu FUNCTION starts and the Sub Display displays the items. Select the item using the FUNCTION key. ②...
Operation Method The Independent Overheating Prevention Device Select the item in function menu ① Set the temperature using the "▼▲". ℃ 220 Notes for the independent overheating prevention device In case there is a small difference between the set values of temperature for the independent overheating prevention device and that of controller, the independent overheating prevention device may be activated and stops the operation.
Operation Method Temperature Rise/Fall Data (Reference Data) The temperature rise/fall data show the reference value when the unit is operated without load. The program should be created based on the data collected by operating the unit with samples put there. Temperature rise data DNE810 ℃...
Handling Precautions WARNING! If a problem occurs If smoke or strange odor should come out of this unit for some reason, turn off the power key right away, and then turn off the circuit breaker and the main power. Immediately contact a service technician for inspection.
Maintenance Method Daily Inspection and Maintenance WARNING! Disconnect the power cable from the power source when doing an inspection or maintenance unless needed. Perform the daily inspection and maintenance after returning the temperature of this unit to the normal one. Do not disassemble this unit.
Long storage and disposal When not using this unit for long term / When disposing CAUTION! When not using this unit for long term… Turn off the power and disconnect the power cord. WARNING! When disposing… Keep out of reach of children. Remove the door and driving parts.
In the Event of Failure… Safety Device and Error Code This unit has an automatic diagnosis function built in the controller and safety devices independent of the controller. The following table shows the purpose, operation and corrective actions against error of safety devices.
Page 62
In the Event of Failure… Safety Device and Error Code Safety Device Purpose State Display Cause/Solution TROUBLE lamp blinks Failure in main relay. To notice Main relay impossibility of Alarm buzzer Check the cause by trouble heater circuit shut activated contacting detection Main relay Er...
In the Event of Failure… Trouble Shooting Refer to page 58 "Safety Device and Error Code" for the state of unit at activation of safety device and corrective actions. Problem Possible Cause Solution Check power connection and The sub indicator does not Power is not supplied.
In Case of Request for Repair If the failure occurs, stop the operation, turn OFF the power switch, and unplug the power plug. Please contact the sales agency that this unit was purchased, or the Yamato Scientific's sales office. < Check following items before contact >...
Specification DNE400/410 DNE600/610 DNE810 DNE910 Method Forced circulation Temperature control Room temp.+ 5℃ to 210℃ (Ambient temp.: 23℃, No load, Damper: full opened) range Temperature ±0.5℃ (at 210℃) adjustment accuracy Temperature ±2.0℃ (at 210℃) distribution accuracy Time required to reach Approx.
Page 66
Specification DNE400/410 DNE600/610 DNE810 DNE910 Capacity 150L 300L 540L Withstand load 15kg/one shelf of shelf Number of shelf bracket step Interval of 30mm shelf bracket steps DNE400: DNE600: 115V AC, 10A 115V AC, 12A Power supply 220V AC, 11.5A 220V AC, 14A DNE410: DNE610: (50/60Hz)
Page 69
Wiring Diagram DNE910 AC220V CONT Symbol Part name Symbol Part name Earth leakage breaker Triac FM1/FM2 Fan motor Independent overheating prevention device H1/H2 Heater Current transformer TM1/TM2 Terminal block TH1/TH2 Temperature sensor (K thermocouple) TF1/TF2 Transformer CONT PLANAR board Relay (heater) Display circuit board Relay (fan)
Reference List of Dangerous Substances Never use explosive substances, flammable substances and substances that include explosive or flammable ingredients in this unit. EXPLOSIVE Ethylene glycol dinitrate (nitro glycol), Glycerin trinitrate (nitroglycerine), Cellulose nitrate (nitrocellulose), and other explosive nitrate esters Trinitrobenzene, Trinitrotoluene, Trinitrophenol (picric acid), and other explosive EXPLOSIVE: nitro compounds Acetyl hidroperoxide (peracetic acid), Methyl ethyl ketone peroxide, Benzyl...
Page 73
Responsibility Please follow the instructions in this document when using this unit. Yamato Scientific has no responsibility for the accidents or breakdown of device if it is used with a failure to comply. Never conduct what this document forbids. Unexpected accidents or breakdown may result in.