Download Print this page

Otto Bock C-Leg 3C98-3 Instructions For Use Manual page 25

Implant system
Hide thumbs Also See for C-Leg 3C98-3:

Advertisement

Available languages

Available languages

CAUTION
Overloading due to changes in body weight when carrying heavy objects, backpacks or
children
>
Falling due to unexpected behaviour of the product.
>
Falling due to breakage of load-bearing components.
>
Skin irritation due to defects on the hydraulic unit with leakage of liquid.
Inform the patient that the behaviour of the product can change due to increased weight. It
is possible that the swing phase may not be triggered, or triggered at the wrong time.
Inform the patient that the maximum permissible body weight must not be exceeded due to
the additional weight.
CAUTION
Using the product while switched off
Falling due to unexpected behaviour of the product because of changed damping behaviour.
Before using the product, switch it on by connecting the power supply and battery charger.
3.8 Notes on the safety modes
CAUTION
Using the product in safety mode
Falling due to unexpected product behaviour because of changed damping behaviour.
The warnings/error signals (Warnings/error signals) have to be observed.
Particular caution is necessary when using a bicycle without a freewheel (with a fixed gear).
CAUTION
Safety mode cannot be activated due to malfunction caused by water penetration or
mechanical damage
Falling due to unexpected product behaviour because of changed damping behaviour.
Using the product when it is defective is prohibited.
The product must be inspected by an authorised Ottobock Service Centre.
CAUTION
Safety mode cannot be deactivated
Falling due to unexpected product behaviour because of changed damping behaviour.
If safety mode cannot be deactivated by recharging the battery, a permanent error has
occurred.
Using the product when it is defective is prohibited.
The product must be inspected by an authorised Ottobock Service Centre.
CAUTION
Safety signal occurs (ongoing vibration)
Falling due to unexpected product behaviour because of changed damping behaviour.
The warnings/error signals (Warnings/error signals) have to be observed.
After the safety signal has been emitted, further use of the product is prohibited.
The product must be inspected by an authorised Ottobock Service Centre.
25

Advertisement

loading

This manual is also suitable for:

C-leg 3c88-3