Adjustments; Set To Default Values - HP 16530A Service Manual

Digitizing oscilloscope module
Table of Contents

Advertisement

HP 16530/31 A- Adjustments
4-13. ADJUSTMENTS
Adjustments In this section must be done in their entirety and in the sequence listed in table 4-2.
Table 4-2. Sequence of Adjustments
SEQUENCE
ADJUSTMENT
1
Set to Default Values
2
DC Balance
3
DC Matching
4
Attenuator Compensation
5
HF Compensation
4-14. Set to Default Values
This will set the calibration factors to known nominal values.
PROCEDURE:
a. Turn power on.
PARAGRAPH
4-14
4-15
4-16
4-17
4-18
b. From the startup screen in figure 4-22, touch these fields in the sequence below:
1. System
2. Oscilloscope
3. Channel
4. Calibration
5. Cal Choice
6. Set to Default
4-16
NO
;

Advertisement

Table of Contents
loading